Product Manual
SE-1200 Series Electrocardiograph User Manual Technical Specifications
- 155 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1, IEC/EN
60601-2-25, ANSI/AAMI EC11, IEC/EN 60601-2-51, UL
60601-1, CAN/CSA C22.2 No.601.1
Anti-electric-shock type: Class І with internal power supply
Anti-electric-shock degree: CF type with defibrillation-proof
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of
application in the presence of
flammable gas:
Equipment not suitable for use in the presence of flammable
gas
Working mode: Continuous operation
EMC: Group І, Class A
NC <10μA (AC) / <10μA (DC)
Patient
Leakage
Current:
SFC <50μA (AC) / <50μA (DC)
NC <10μA (AC) / <10μA (DC)
Patient
Auxiliary
Current:
SFC <50μA (AC) / <50μA (DC)










