SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
- 146 -
3. Clean them in warm water and make sure there is no remainder gel.
4. Dry the electrodes with a clean dry cloth or air dry naturally.
CAUTION
Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
14.1.3 Cleaning the Print Head
To clean the print head:
1. Open the recorder casing and remove the paper.
2. Wipe the print head gently with a clean soft cloth damped in 75
% alcohol.
For stubborn stain, soak it with a little alcohol first and wipe it off with a clean soft cloth.
3. After air drying, load the recorder paper and shut the recorder casing.
Dirty and soiled thermal print head will deteriorate the printing definition. So it should be cleaned
at least once a month regularly.
CAUTION
Prevent the detergent from seeping into the main unit while cleaning. Do not immerse the
unit or the patient cable into liquid under any circumstances.
14.2 Disinfection
Disinfection of the main unit is not necessary need in daily maintenance, it is only
necessary in operating room. In that case, please use hospital standard disinfectant.
NOTE: Clean and disinfect the chest and limb electrodes after each use.
CAUTION
1. Do not use high-temperature, high-pressure vapour or ionizing radiation as
disinfection methods.
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.