SE-1200 Series Electrocardiograph User Manual System Setup
- 142 -
11.11.5 Setting External Output
Press Tab or Shift + Tab to move the cursor to the External Output list box on the Other Setup
Window, and then press the Up or Down arrow to highlight an option.
The extern output socket is equipped in the electrocardiograph, through which the
electrocardiograph can send signals to the external equipment.
If External Input is set to On, and External Output is set to Standard or Triggered, the
electrocardiograph sends the signals which it receives.
If External Input is set to Off, and External Output is set to Standard, the electrocardiograph
sends ECG signals of rhythm lead I.
If External Input is set to Off, and External Output is set to Triggered, the electrocardiograph
sends pulses with the height of 4V and the width of 40ms, based on the data of rhythm lead I.
11.11.6 Setting Pacemaker Detection Sensitivity
Press Tab or Shift + Tab to move the cursor to Pacemaker Detection Sens. on the Other Setup
Window, and then press the Up or Down arrow to highlight Low or High.
When Pacemaker Detection Sens. is set to High, the pacemaker signals are easy to be detected.
When Pacemaker Detection Sens. is set to Low, the pacemaker signals are not easy to be
detected.
For more information about the detectable parameter ranges, please refer to Appendix 1,
“Technical Specifications”.
NOTE: If you select Pacemaker in the Patient Information Setup window, the pacemaker
sensitivity can only be set in the Patient Information window, and the setting of
Pacemaker Detection Sens. on the Other Setup Window is ineffective.