Product Manual
SE-1200 Series Electrocardiograph User Manual System Setup
- 124 -
To Set Local IP
For the cross-network transmission, set the first two sections of the Local IP item to the first two
sections of the IP of the PC. Set the third section of the Local IP item to the network segment of
the electrocardiograph which depends on the configuration of Router. The last section of the
Local IP item can be set at random.
For the same network transmission, set the first three sections of the Local IP item to the first
three sections of the IP of the PC. The last section of the Local IP item can be set at random, but
it can’t be the same as the last section of the IP of the PC.
To Set Gateway
Set the first three sections of the Gateway item to the first three sections of the IP of the PC. The
last section of the Gateway item must be set to 1.
To Set Subnet Mask
You can set Subnet Mask in the Transmission Setup window. Commonly, Subnet Mask is set to
255.255.255.000.
11.6 Lead Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Lead on the System Setup screen, press
to open the Lead Setup
window.