SE-1200 Series Electrocardiograph User Manual System Setup
- 119 -
You can set ID Mode to Auto, Time or Manual.
When ID Mode is set to Auto, the patient ID can be automatically generated. The patient ID
range is 0~1999, 999, 999.
When ID Mode is set to Time, the patient ID can be automatically generated according to the
time when you press the START/STOP key to print an ECG report.
When ID Mode is set to Manual, you can enter the patient ID manually in the Patient
Information window. (Only 30 ASCII characters can be input.)
11.4.3 Selecting ID Hint
Press Tab or Shift + Tab to move the cursor to ID Hint in the Patient Information Setup window,
and then press the Up or Down arrow to select On or Off.
In the auto, rhythm or off mode, when ID Mode is set to Manual and ID Hint is set to On, if
you do not input the patient ID before pressing the START/STOP key, a hint will pop up to
remind you to input the patient ID.
11.4.4 Setting Prompt
Press Tab or Shift + Tab to move the cursor to Prompt in the Patient Information Setup window,
and then press the Up or Down arrow to select Confirmed By or Unconfirmed.
When Prompt is set to Confirmed By, if Physician is selected and the physician name is input
in the Patient Information window, the physician’s name is printed in the ECG reports.
When Prompt is set to Unconfirmed, Unconfirmed Report is printed in the ECG reports.
11.4.5 Selecting H/W Unit
Press Tab or Shift + Tab to move the cursor to H/W Unit in the Patient Information Setup
window, and then press the Up or Down arrow to select cm/kg or inch/lb..
11.4.6 Selecting BP Unit
Press Tab or Shift + Tab to move the cursor to BP Unit in the Patient Information Setup window,
and then press the Up or Down arrow to select mmHg or kpa.