When Record Sequence is set to Synchronous, all leads are printed simultaneously. The start
time of each group is the same.
11.3.1.4 Setting Gain
Press Tab or Shift + Tab to move the cursor to the Gain list box on the Setup 1 tab, and then
press the Up or Down arrow to highlight an option.
You can set the indicated height of 1mV ECG on the paper.
You can set Gain to 10mm/mV, 20mm/mV, 10/5mm/mV, AGC, 2.5mm/mV or 5mm/mV.
AGC means auto gain control. When ECG signals vary greatly, AGC can be selected to adjust
the gain automatically according to actual signals.
10/5mm/mV means that the gain of limb leads is set to 10mm/mV, while the gain of chest leads
is set to 5mm/mV.
11.3.1.5 Selecting Recording Device
Press Tab or Shift + Tab to move the cursor to the Record Device list box on the Setup 1 tab.
Press the Up or Down arrow to select an option.
You can set Record Device to Thermal, HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668,
HP-DJ-D4368 or HP-LJ-P2035.
HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668, HP-DJ-D4368 and HP-LJ-P2035 are the types
of the USB printer, when HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668, HP-DJ-D4368 or
HP-LJ-P2035 is selected, you should connect the corresponding USB printer to USB socket 2 of
the electrocardiograph with a special cable. Press the START/STOP key to print ECG reports by
using the USB printer.
WARNING
If the printer used is not the type listed above, additional safety measures (such as
applying an isolation transformer to supply the medical system) should be taken when the
safety of the medical system has not been evaluated. If in doubt, consult our technical
service department or your local distributor