Product Manual

SE-1200 Series Electrocardiograph User Manual System Setup
- 113 -
In the auto mode, when Record Style is set to 3×4, ECG waves of 12 leads are printed in 4
groups of 3.
In the auto mode, when Record Style is set to 3×4+1R, ECG waves of 12 leads are printed in 4
groups of 3, with the ECG wave of one rhythm lead on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 3×4+3R, ECG waves of 12 leads are printed in 4
groups of 3, with the ECG waves of three rhythm leads on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 6×2, ECG waves of 12 leads are printed in 2
groups of 6.
In the auto mode, when Record Style is set to 6×2+1R, ECG waves of 12 leads are printed in 2
groups of 6, with the ECG wave of one rhythm lead on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 12×1, ECG waves of 12 leads are printed
simultaneously.
11.3.1.2 Selecting Recording Mode
Press Tab or Shift + Tab to move the cursor to the Record Mode list box on the Setup 1 tab, and
then press the Up or Down arrow to highlight Save Paper or Quickly.
When Record Mode is set to Save Paper, 10s after pressing the START/STOP key on the main
screen, an ECG report is printed. The patient information, measure information, interpretation
and ECG waves are printed on the same paper.
When Record Mode is set to Quickly, pressing the START/STOP key on the main screen to
begin printing an ECG report immediately, the patient information, measure information,
interpretation and ECG waves are printed on two pieces of paper.
NOTE:
1. In the auto mode, only when Sampling Mode is set to Real-time Sample, Quickly is
available.
2. When Record Style is set to 3×4, 3×4+1R or 3×4+3R, only Save Paper is available.
When Record Style is set to 12×1, only Quickly is available.
11.3.1.3 Selecting Recording Sequence
Press Tab or Shift + Tab to move the cursor to the Record Sequence list box on the Setup 1 tab,
and then press the Up or Down arrow to highlight Sequential or Synchronous.
When Record Sequence is set to Sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the previous lead group.