Product Manual

SE-1200 Series Electrocardiograph User Manual System Setup
- 110 -
Figure 11-3 Filter Setup window
After setup, press to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press to confirm. Then the System Setup screen appears.
11.2.1 Setting AC Filter
Press Tab or Shift + Tab to move the cursor to the AC Filter list box in the Filter Setup window.
Press the Up or Down arrow to highlight Off or On.
AC Filter suppresses AC interference without attenuating or distorting the ECG signals. Select
On to turn on the function and select Off to turn it off.
NOTE: AC frequency can be set to 50Hz or 60Hz on the Advanced Setup screen
according to local mains supply specifications.
11.2.2 Setting DFT Filter
Press Tab or Shift + Tab to move the cursor to the DFT Filter list box in the Filter Setup window.
Press the Up or Down arrow to highlight an option.