Product Manual
SE-1200 Series Electrocardiograph User Manual System Setup
- 109 -
11.1.6 Setting Interval and Duration
Press Tab or Shift + Tab to move the cursor to the Duration and Interval textboxes in the Work
Mode Setup window, and then input the duration and the interval manually.
In the auto mode, when Sampling Mode is set to Periodic Sample, if Interval is set to 2 min,
Duration is set to 24 min, after pressing the START/STOP key, the printing will be performed
every two minutes and come to 12 times.
11.1.7 Selecting Preview Function
Press Tab or Shift + Tab to move the cursor to Preview in the Work Mode Setup window, and
then press the Up or Down arrow to highlight Off or On.
When Preview is set to On, the preview screen will appear after you press the START/STOP
key in the auto or off mode.
When Preview is set to Off, the preview screen will not appear after you press the
START/STOP key.
11.1.8 Selecting Auto Extend Record
Press Tab or Shift + Tab to move the cursor to Auto Extend Record in the Work Mode Setup
window, and then press the Up or Down arrow to highlight Off or On.
When Auto Extend Record is set to On, if arrhythmia is detected in the auto or off mode, a hint
will pop up to ask you whether to print an extra rhythm report after the 12-lead ECG report.
11.2 Filter Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Filter on the System Setup screen, press to open the Filter Setup
window.
In the Filter Setup window, you can set AC Filter, EMG Filter, DFT Filter, and Lowpass
Filter.