Product Manual

SE-1200 Series Electrocardiograph User Manual System Setup
- 108 -
When the auto display style is set to 3×4+1R, ECG waves of 12 leads are displayed in 4 groups
of 3, with the ECG wave of one rhythm lead on the bottom.
When the auto display style is set to 3×4+3R, ECG waves of 12 leads are displayed in 4 groups
of 3, with ECG waves of three rhythm leads on the bottom.
When the auto display style is set to 6×2, ECG waves of 12 leads are displayed in 2 groups of 6.
When the auto display style is set to 6×2+1R, ECG waves of 12 leads are displayed in 2 groups
of 6, with the ECG wave of one rhythm lead on the bottom.
When the auto display style is set to 12×1, ECG waves of 12 leads are displayed on one screen
simultaneously.
11.1.5 Selecting Sampling Mode
Press Tab or Shift + Tab to move the cursor to the Sampling Mode list box in the Work Mode
Setup window, and then press the Up or Down arrow to highlight an option.
When Sampling Mode is set to Pre-Sample, 10s ECG data sampled before pressing the
START/STOP key will be printed out.
NOTE: When Sampling Mode is set to Pre-Sample, if you press the START/STOP key
before the electrocardiograph samples for 10s, the recorder will not respond.
When Sampling Mode is set to Real-time Sample, 10s ECG data sampled after pressing the
START/STOP key will be printed out.
When Sampling Mode is set to Triggered Sample, after pressing the START/STOP key, if
Arrhythmia ECG data, including Asystole, Ventricular Fibrillation/Ventricular Tachycardia,
5>PVCS>=3, Paired PVCS, Bigeminy, Trigeminy, R ON T, single PVC and Missed Beat, is
detected during the learning course, the printing will be triggered automatically.
When Sampling Mode is set to Periodic Sample, first you should set Duration and Interval.
For example, if Interval is set to 2 min, and Duration is set to 24 min, after pressing the
START/STOP key, the printing will be performed every two minutes and come to 12 times.
NOTE:
1. There is no saving, exporting or transmitting in the Periodic Sample mode.
2. ID and patient information will not be changed while carrying out the periodic printing.