Product Manual

SE-1010 PC ECG User Manual Preparations Before Operation
- 31 -
FRANK lead system is usually adopted when PC ECG is used to produce VCG. The conventional
letter designations for the electrodes and their respective positions are:
E/C2: at the front midline
M/C5: at the back midline
I/C1: at the right mid-axillary line
A/C4: at the left mid-axillary line
C/C3: at 45ยบ angle between the front midline and the left mid-axillary line
F: on the left leg
N: on the right leg
H: on the back of the neck
The first five electrodes (E, M, I, A and C) are all located at the same transverse level --
approximately at the interspace between the fourth rib and the fifth rib.
5.4.3 Attaching Electrodes to the Patient
For Limb Electrodes:
1. Clean the electrode area which is in a short distance above the ankle or the wrist with
alcohol.
2. Daub the electrode area on the limb with gel evenly.
3. Place a small amount of gel on the metal part of the limb electrode clamp.
4. Connect the electrode to the limb, and make sure that the metal part is placed on the
electrode area above the ankle or the wrist.
5. Attach all limb electrodes in the same way.
For Chest Electrodes:
Reed
Connecting to a Lead Wire
Clamp
Suction Bulb
Connecting to a Lead Wire
Cup