Product Manual

SE-1010 PC ECG User Manual Preparations Before Operation
- 29 -
5.3 Connecting the Electrodes of Wireless System
The patient cable includes main cable and lead wires which can be connected to electrodes
according to the colors and identifiers.
Patient Cable of Wireless System
1. Connect the patient cable to DX12 transmitter of wireless system. For details, see Section 3.2,
“Assembling Wireless System”.
2. Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the
corresponding electrodes according to the colors and identifiers. Firmly attach them.
5.4 Attaching Electrodes (for Resting ECG)
The identifiers and color codes of electrodes used comply with IEC/EN requirements. In order to
avoid incorrect connections, the electrode identifiers and color codes are specified in Table 5-1.
Moreover the equivalent codes according to American requirements are given in Table 5-1 too.
Table 5-1 Electrodes and Their Identifiers and Color Codes
European American
WILSON FRANK Identifier Color Code Identifier Color Code
Right arm Right arm R Red RA White
Left arm Left arm L Yellow LA Black
Right leg Right leg N or RF Black RL Green
Left leg Left leg F Green LL Red
Chest 1 I C1 White/Red V1 Brown/Red
Chest 2 E C2 White/Yellow V2 Brown/Yellow
Chest 3 C C3 White/Green V3 Brown/Green
Chest 4 A C4 White/Brown V4 Brown/Blue
Chest 5 M C5 White/Black V5 Brown/Orange
Chest 6 H C6 White/Violet V6 Brown/Violet
Main Cable
Lead Wires
Connecting to the
ECG Sampling Box
Connecting to Electrodes