About this Manual P/N: 01.54.106666-20 Release Date: June 2012 © Copyright EDAN INSTRUMENTS, INC. 2008-2012. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A WARNING label advises against certain actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides useful information regarding a function or a procedure.
Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use........................................................................................................................ 1 1.2 Warnings and Cautions........................................................................................................ 1 1.2.1 General Warnings ................................................................
.6 Inspection Before Test....................................................................................................... 35 5.7 Setting DX12 Transmitter (for Wireless System) ............................................................. 36 5.7.1 Keyboard Locking/Unlocking................................................................................. 38 5.7.2 Menu Settings .........................................................................................................
.6.5 Analyzing HRV ....................................................................................................... 70 6.6.5.1 Editing the HRV Data on the Analysis Screen.............................................. 71 6.6.5.2 Editing the HRV Waveform in the Waveform Window ................................ 72 6.6.5.3 Previewing HRV ........................................................................................... 73 6.6.6 Analyzing HRT.......................................................
.7 Recovery Phase ................................................................................................................. 98 7.8 Exiting the Exercise Test................................................................................................... 98 7.9 About Analysis Screen ...................................................................................................... 99 7.9.1 About Summary Screen......................................................................................
9.2.6 Setting Anti-aliasing.............................................................................................. 123 9.2.7 Selecting QRS Voice ............................................................................................. 123 9.2.8 Selecting Sequence Mode When Sampling .......................................................... 123 9.3 Device Setup ................................................................................................................... 124 9.3.
Chapter 10 Hint Information.................................................................................................... 144 Chapter 11 Cleaning, Care and Maintenance ......................................................................... 146 11.1 Cleaning and Maintaining the Treadmill....................................................................... 146 11.2 Cleaning and Maintaining the Patient Cable and Reusable Electrodes ........................ 146 11.3 Disinfection ......................
SE-1010 PC ECG User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of SE-1010 PC ECG. 1.1 Intended Use SE-1010 PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. SE-1010 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals.
SE-1010 PC ECG User Manual Safety Guidance WARNING 3. EXPLOSION HAZARD - Do not use the system in the presence of flammable anesthetic mixtures with oxygen or other flammable agents. 4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet. Never try to adapt the three-prong plug to fit a two-slot outlet. 5. Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed.
SE-1010 PC ECG User Manual Safety Guidance WARNING 16. SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied as a part of the system, to the multiple portable socket-outlet supplying the system. 17. Do not connect any equipment or accessories that are not approved by the manufacturer or that are not IEC/EN 60601-1-1 approved to the system.
SE-1010 PC ECG User Manual Safety Guidance WARNING 25. You should purchase computer, printer, treadmill, ergometer, BP monitor and bar code reader from the manufacturer. Otherwise, the manufacturer will not be held responsible for the maintenance of the PC hardware, operating system and other accessories. 26. If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel. 27.
SE-1010 PC ECG User Manual Safety Guidance 1.2.3 General Cautions CAUTION 1. Avoid liquid splash and excessive temperature. The temperature must be kept between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and 55 ºC during transportation and storage. 2. Do not use the equipment in a dusty environment with bad ventilation or in the presence of corrosive. 3.
SE-1010 PC ECG User Manual Safety Guidance 1.2.5 Preparation and Operation Warnings (for Exercise ECG) WARNING 1. Test the safety stop (mushroom type) and safety stop (cord type) of the treadmill before using the system. 2. During the exercise test, ensure that there are at least 2 experienced physicians present. One of them observes the patient and deals with the emergency. 3.
SE-1010 PC ECG User Manual Safety Guidance 1.2.6 Contraindications (for Exercise ECG) Absolute Contraindications: 1. Acute MI (within 2 days) 2. High-risk unstable angina 3. Hemodynamic compromise caused by uncontrolled cardiac arrhythmia 4. Symptomatic severe aortic stenosis 5. Heart failure with clinic episode uncontrolled 6. Acute pulmonary embolus or pulmonary infarction 7. Acute myocarditis or pericarditis 8. The patient opposes the test. Relative Contraindications: 1. Left main coronary stenosis 2.
SE-1010 PC ECG User Manual Safety Guidance 1.3 List of Symbols Equipment or part of CF type with defibrillator proof Caution Consult Instructions for Use Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
SE-1010 PC ECG User Manual Safety Guidance Class Ⅱ Transmission Status Indicator of Bluetooth Burglar Lock Power Supply Indicator of DX12 Receiver -9-
SE-1010 PC ECG User Manual Introduction Chapter 2 Introduction SE-1010 PC ECG has similar functions with an ordinary electrocardiograph. ECG data can be sampled, analyzed and stored in a PC, and it can be saved in PDF, Word, BMP or JPG format. ECG waves can be frozen and reviewed. Auto measurement and diagnosis are available, and the diagnosis template can be edited. SE-1010 PC ECG can also be invocated by Smart ECG Net version 1.3 or above.
SE-1010 PC ECG User Manual Introduction Wired System of SE-1010 PC ECG System 1 Resting ECG of Wired System Patient Patient Cable Resting PC (Manually Configured) ECG Cable DP12 ECG Sampling Box USB Cable Printer (Manually Configured) 2 Exercise ECG of Wired System Treadmill or Ergometer (Manually Configured) Patient Serial Cable Patient Cable Exercise PC (Manually Configured) ECG Cable DP12 ECG Sampling Box USB Cable Printer (Manually Configured) - 11 - Exercise ECG Cable BP Monitor (Ma
SE-1010 PC ECG User Manual Introduction Wireless System of SE-1010 PC ECG System 1 Resting ECG of Wireless System Patient Patient Cable Resting PC (Manually Configured) ECG Cable DX12 Receiver DX12 Transmitter USB Cable Printer (Manually Configured) 2 Exercises ECG of Wireless System Treadmill or Ergometer (Manually Configured) Patient Serial Cable Patient Cable Exercise PC (Manually Configured) ECG Cable DX12 Receiver DX12 Transmitter Exercise USB Cable Printer (Manually Configured) ECG
SE-1010 PC ECG User Manual Introduction 2.2 DP12 ECG Sampling Box of Wired System DP12 ECG Sampling Box Appearance Front Panel Lamp USB Socket Name Explanation Lamp When the ECG sampling box is powered by the PC, the lamp will be lit.
SE-1010 PC ECG User Manual Introduction USB Socket Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 6 GND 2 VCC 7 GND 3 QRS 8 GND 4 GND 9 D- 5 GND 10 D+ WARNING 1. When the computer connected to the USB cable is powered on, do not connect the USB cable to the DP12 ECG sampling box; when the system is powered on, do not disconnect the USB cable from the ECG sampling box. 2.
SE-1010 PC ECG User Manual Introduction Patient Cable Socket Definitions of corresponding pins: Pin Signal Pin Signal Pin Signal 1 C2 / V2 6 SH 11 F / LL 2 C3 / V3 7 NC 12 C1 / V1 or NC 3 C4 / V4 8 NC 13 C1 / V1 4 C5 / V5 9 R / RA 14 RF (N) /RL or NC 5 C6 / V6 10 L / LA 15 RF (N) / RL NOTE: The left side of “/” is European standard, and the right side is American standard.
SE-1010 PC ECG User Manual Introduction WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall comply with the valid version of the standard IEC/EN 60601-1-1.
SE-1010 PC ECG User Manual Introduction Front Panel Patient Cable Socket Patient Cable Socket :Applied part of type CF with defibrillator proof :Caution Definitions of corresponding pins: Pin Signal Pin Signal Pin Signal Pin Signal 1 NC 6 C5/V5 11 NC 16 L/LA 2 F/LL 7 NC 12 C2/V2 17 NC 3 NC 8 C4/V4 13 NC 18 R/RA 4 C6/V6 9 NC 14 C1/V1 19 NC 5 NC 10 C3/V3 15 NC 20 N/RL NOTE: The left side of “/” is European standard, and the right side is American standard.
SE-1010 PC ECG User Manual Introduction DX12 Receiver Appearance USB Socket USB Socket - 18 -
SE-1010 PC ECG User Manual Introduction Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 5 GND 2 VCC 6 GND 3 QRS 7 GND 4 GND 8 D- 5 GND 10 D+ Lock 2.
SE-1010 PC ECG User Manual Introduction ♦ Reports can be printed in PDF, Word, JPG or BMP format ♦ Supporting multi-language ♦ Updated to be a network electrocardiograph, transmitting ECG data over LAN or WAN or INTERNET ♦ Automatic baseline adjustment for optimal printing ♦ High performance filters guarantee stable ECG waveforms ♦ Real-time analysis, real-time displaying and printing 12-lead simultaneous ECG waveforms ♦ Nine analysis functions including Normal ECG, Frequency ECG, High Frequency ECG, QT D
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System Chapter 3 Assembling SE-1010 PC ECG System 3.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System For Resting ECG of wired system, 1. Insert plug 1 of the patient cable into socket 3 of DP12 ECG sampling box. 2. Insert plug 8 of the cable into socket 4 of DP12 ECG sampling box. 3. Insert plug 9 of the cable into the USB socket of the PC. 4. Connect a printer to the PC. 5. Insert the Sentinel into the USB socket of the PC. 6.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 3.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 12 11 Burglar Lock Assembly Drawing For Resting ECG of wireless system, 1. Insert plug 1 of the patient cable into socket 2 of DX12 transmitter. 2. Insert plug 4 of the cable into socket 6 of DX12 receiver. 3. Insert plug 5 of the cable into the USB socket of the PC. 4. Connect a printer to the PC. 5. Insert the Sentinel into the USB socket of the PC. 6.
SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software Chapter 4 Installing SE-1010 PC ECG Software 4.1 System Running Environment 4.1.
SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software 4.2 About Installation Window Insert the installation CD into CD-ROM, and double-click on Setup.exe to open the following installation window. Figure 4-1 Installation Window Click on the Install button to install PC ECG. Click on the Next button continually during installation. After installing PC ECG, click on the Install button in the installation window. Then the Environment Detection window pops up.
SE-1010 PC ECG User Manual Preparations Before Operation Chapter 5 Preparations Before Operation 5.1 Preparing the Patient 5.1.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area where others can’t see the patient.
SE-1010 PC ECG User Manual Preparations Before Operation 5.2 Connecting the Electrodes of Wired System WARNING The performance and electric shock protection can be guaranteed only if the original patient cable and electrodes of the manufacturer are used. The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers.
SE-1010 PC ECG User Manual Preparations Before Operation 5.3 Connecting the Electrodes of Wireless System The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers. Connecting to the ECG Sampling Box Connecting to Electrodes Main Cable Lead Wires Patient Cable of Wireless System 1. Connect the patient cable to DX12 transmitter of wireless system. For details, see Section 3.2, “Assembling Wireless System”. 2.
SE-1010 PC ECG User Manual Preparations Before Operation 5.4.1 Wilson Lead System C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the left midclavicular line C5: Left anterior axillary line at the horizontal level of C4 C6: Left midaxillary line at the horizontal level of C4 5.4.
SE-1010 PC ECG User Manual Preparations Before Operation FRANK lead system is usually adopted when PC ECG is used to produce VCG.
SE-1010 PC ECG User Manual Preparations Before Operation 1. Clean the electrode area on the chest surface with alcohol. 2. Daub the round area of 25mm in diameter on each electrode area with gel evenly. 3. Place a small amount of gel on the brim of the chest electrode’s metal cup. 4. Place the electrode on the chest electrode area and squeeze the suction bulb. Unclench it and the electrode is adsorbed on the chest. 5. Attach all chest electrodes in the same way.
SE-1010 PC ECG User Manual Preparations Before Operation 5.5 Attaching Electrodes to the Patient (for Exercise ECG) The identifiers and color codes of electrodes used comply with IEC/EN requirements. In order to avoid incorrect connections, the electrode identifiers and color codes are specified in Table 5-2. Moreover the equivalent codes according to American requirements are given in Table 5-2 too.
SE-1010 PC ECG User Manual Preparations Before Operation The Precordial Electrodes’ Positions on Body Surface: C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the left midclavicular line C5: Left anterior axillary line at the horizontal level of C4 C6: Left midaxillary line at the horizontal level of C4 The Extremity Electrodes’ Positions on Body Surface: R/L: below
SE-1010 PC ECG User Manual Preparations Before Operation 5.6 Inspection Before Test In order to avoid safety hazards and get good ECG records, the following inspection procedure is recommended before operation. 1. Environment: ♦ Make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imaging equipment etc. Switch off these devices when necessary.
SE-1010 PC ECG User Manual Preparations Before Operation 5.7 Setting DX12 Transmitter (for Wireless System) Switch on DX12 receiver and install batteries on DX12 transmitter. Press to start up DX12 transmitter, and then the company information and the main screen will be displayed. Figure 5-1 Main Screen When the main screen is displayed, press to switch the leads. is not displayed on the main screen, you have to match the If the Bluetooth connection icon device manually.
SE-1010 PC ECG User Manual 1. Press Preparations Before Operation to enter the menu screen. Figure 5-2 Menu Screen 2. Press to display Match Device item in black, and then press with a prompt “Inquiring…”.
SE-1010 PC ECG User Manual 3. Preparations Before Operation When the receiver is found, the address of DX12 receiver will be displayed on the screen in 10 seconds. Press to start up the Bluetooth connection. The Bluetooth connection will be displayed on the main screen of DX12 transmitter and the Bluetooth status indicator illuminates in kelly when DX12 transmitter and DX12 receiver are matched successfully.
SE-1010 PC ECG User Manual Preparations Before Operation 5.7.2 Menu Settings Press on the main screen to enter the menu screen (Figure 5-2). Press screen to display an item in black, and then press on the menu to enter the setting screen of this item. Table 5-3 Menu Menu Back Light Auto Sleep Option Description On Select On to turn on the backlight of LCD screen. Off Select Off to turn off the backlight of LCD screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Chapter 6 Operation Instructions for Resting ECG Double-click on the shortcut icon the desktop icon for SE-1010 PC ECG. on the desktop to display the main screen. is NOTE: Do not use other software when using PC ECG software. Figure 6-1 Toolbar of Main Screen The toolbar contains six buttons. From left to right, they are New Patient, STAT ECG, Data Manager, System Setting, Lead Placement and Exit.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can set the unit, print, frequency of AC and the saving path of source data based on your needs. Click on the OK button after setup, the system will enter the main screen automatically. NOTE: You should install the software to the saving path of source data after the uninstallation and reinstallation; otherwise, the software needs a new configuration. 6.1 Viewing Lead Placement Information 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.2 Selecting a Patient Record to Start a New Test Click on the Data Manager button on the main screen (Figure 6-1) to open the Data Manager screen (Figure 6-3). Figure 6-3 Data Manager Screen 1. Select a search item in the pull-down list on the Data Manager screen. Then all the patient records which meet the search condition are listed in the patient information list. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 3. Or, click on Advanced Search to display the Search Condition window. Enter the search conditions, and click on the Search button, and all the patient records which meet the conditions will be displayed in the patient information list. 4. Click on the patient record in the patient information list and click on the Select button to open the Patient Information window.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.3 Entering New Patient Information 6.3.1 Entering New Patient Information Manually If the patient is a new one, you can click on the New Patient button on the main screen (Figure 6-1) to display the Patient Information window. Then you need to input the patient’s related information. 1. Enter basic information, such as patient ID, name, sex and age.
SE-1010 PC ECG User Manual 3. Operation Instructions for Resting ECG Enter information of doctor and department 1) Enter information of physician, technician or Req. physician Click on the pull-down list , and then click on the Edit button to display the Edit window. Enter the doctor name in the textbox, and then click on the Add button. The doctor name will be displayed in List.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 2) Enter information of department Click on the pull-down list , and then click on the Edit button to display the Edit Department window. Enter the department name in the textbox, and then click on the Add button. The doctor name will be displayed in List. Meanwhile, you can also enter the department name in the textbox of Dept. item, and then click on the OK button in the Patient Information window.
SE-1010 PC ECG User Manual 4. Operation Instructions for Resting ECG Confirm pacemaker information If you select Pacemaker in the Print Setting window, Pacemaker appears in the Patient Information window. Select Pacemaker to detect very small pacemaker pulses. However, when pacemaker pulse enhancer is on, the system is very sensitive, and should not be close to equipment emitting high frequency radiation. High frequency radiation can interfere with pacemaker pulse detection and normal ECG acquisition.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.4 Selecting Sampling Type You can select a sampling type in the Patient information window. 6.5 Sampling Resting ECG After inputting the patient information, click on the OK button in the Patient information window to open the ECG sampling screen. Before sampling, if you do not connect the PC to the ECG sampling box, the following hint will pop up.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The system begins to pre-sample ECG. Figure 6-5 Pre-Sampling Screen 6.5.1 Specifying Display Mode There are eight display modes including 12*1, 6*2, 3*4, 6*1, 3*1, 3*4+1, 3*4+3 and 6*2+1. When the display mode is set to 12*1, 12-channel ECG waves are displayed on one screen simultaneously. When the display mode is set to 6*2, 12-channel ECG waves are displayed in 2 groups of 6 on one screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG When the display mode is set to 3*1, 3-channel ECG waves are displayed on one screen simultaneously. When the display mode is set to 3*4+1, 12-channel ECG waves are displayed in 4 groups of 3 and one rhythm lead on one screen. When the display mode is set to 3*4+3, 12-channel ECG waves are displayed in 4 groups of 3 and three rhythm leads on one screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.5 Recording ECG Data When the pre-sample ECG waves are steady, you can click on the Start button to save the sampled ECG data to the designated directory. For details, please refer Section 9.1.4, “Specifying the Storage Path of the ECG Data”. Figure 6-6 ECG Sampling Screen NOTE: After you click on the Start button, the system will save the sampled ECG data.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-7 Wave Review Window Click on Exit to return to the ECG sampling screen. NOTE: The display modes in the Wave review window are the modes you select on the ECG sampling screen. 3*1 and 6*1 display modes are displayed in 3*4 and 6*2 modes. 6.5.7 Stopping Sampling Data After clicking on the Start button, there are two ways to stop sampling data. 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.8 Printing ECG Waves Click on the Print button on the ECG sampling screen to print the ECG waves on the ECG sampling screen. NOTE: 1. You can set the printer type on the Printer and Faxes screen. For details, please refer to Section 6.6.10, “Printing ECG Reports”. 2. The report can be printed in white- black or color. The report color is defined by setting the printer type and can be observed on the preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.1 Viewing the Waveform Click on the Waveform tab on the normal ECG analysis screen to open the Waveform window (Figure 6-8). Figure 6-8 Normal ECG - Waveform Window You can choose a speed, a gain and a display mode for the displayed waves. Click on the Measure button in the Waveform window (Figure 6-8). Click on one point on the wave, and then drag the mouse to another point.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The common parameters are displayed in the circle in Figure 6-8. These parameters are as follows. Designation Description Heart Rate Heart Rate P Duration P-wave duration of the current lead PR Dur. P-R interval of the current lead QRS Dur.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 1. Enter your own opinions in the Diagnosis textbox, and then click on the Save button. 2. Or, double-click on the necessary results required to be added in the Glossary textbox, the selected results will be displayed in the Diagnosis textbox, and then click on the Save button. If the Integration function is activated, the data can be uploaded to Smart ECG Net system after you click on the Save button. 6.6.1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG To Edit Waveform on the Analysis Screen You can press a lead button in to display magnified average templates of this lead. When you press more than one lead button, magnified average templates of these leads will be overlapped with the same central axis. When you press ALL, magnified average templates of all leads will be overlapped with the same central axis. You can set the speed and the gain of average templates.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.3 About the Detail Information Window Click on the Detail information tab on the normal ECG analysis screen to open the Detail information window. This window displays lead parameter values as Figure 6-10 shows. Figure 6-10 Normal ECG - Detail Information Window Click on the Export Excel button to export an Excel file. To Edit Diagnosis Result in the Detail Information Window For details, please refer to Section 6.6.1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.4 About the Rhythm Wave Window Click on the Rhythm Wave tab on the normal ECG analysis screen to open the Rhythm wave window. This window displays rhythm wave as Figure 6-11 shows. Figure 6-11 Normal ECG - Rhythm Wave Window You can set the gain, the speed and the lead of the displayed ECG waves. You can click on Previous Page or Next Page to display the waves of the previous or next page.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 4. Click on Close to close the normal ECG preview screen and return to the previous screen. Figure 6-12 ECG Wave NOTE: The report title is in one line by default, you can also change the line to two lines. Find the text named PCECG.ini in the installation path, modify LineofTitle=0 to LineofTitle=1, and then save the text to change the line to two lines.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Report Title in Two Lines 6.6.2 Analyzing QT Dispersion Click on the QTD button to open the QT Dispersion screen. QT Dispersion: The difference between the largest QT interval and the shortest QT interval based on the synchronous 12-lead surface ECG. The QT interval is a measurement of the time difference between the start of the Q wave and the end of the T wave.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.1 Editing Waveform on the QT Dispersion Screen There are red and blue triangle icons on top of the displayed ECG waves. You can click on the triangle icon to change the color. represents the current chosen R wave represents the position of R wave You can set the speed and the gain of the displayed ECG waves. The lead number and the lead type of the displayed ECG waves can also be chosen. 6.6.2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.3 Previewing QT Dispersion Click on the Preview button on the QT Dispersion screen to open the QT Dispersion preview screen. is the toolbar on the QT Dispersion preview screen. 1. Click on Print(P) to print the report. 2. Click on Next Page/Prev Page to switch to the previous/next preview page. 3. Click on Two Page to preview two pages on one screen simultaneously. 4. Click on Zoom In/ Zoom Out to magnify/minify the preview page. 5.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3 Analyzing Frequency ECG Click on the FCG button to open the Frequency ECG screen. Frequency ECG displays characteristic waves of ECG signal spectrum. There are two tabs on the Frequency ECG screen: Two leads compare and 12-lead power spectrum. 6.6.3.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Every parameter of Frequency ECG is shown in the following table. Power Spectrum Phase Shift Designation Definition 1/2 If the first peak value is lower than 90% of the second peak value, it is positive (‘+’). Otherwise it is minus (‘-’). 1-N The first peak is too low or disappears. (The frequency value of the first peak’s position * 60 = heart rate. For example, if the frequency value of the first peak’s position is 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3.2 About 12-lead Power Spectrum Window The 12-lead power spectrum window displays the power spectrum of 12-lead ECG waves. Figure 6-16 Frequency ECG - 12-lead Power Spectrum Window 6.6.3.3 Previewing Frequency ECG Click on the Preview button to open the Frequency ECG preview screen. is the toolbar on the Frequency ECG preview screen. 1. Click on Print(P) to print the report. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-17 Two-lead Comparison Report Figure 6-18 12-lead Power Spectrum Report - 67 -
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.4 Analyzing High Frequency ECG In a conventional electrocardiogram, only these ECG signals, of which the amplitude is in the millivolt range and the frequency is less than 100 Hz, are visible. Those ECG signals, of which the amplitude is in the microvolt range and the frequency is between 150Hz and 250Hz, are invisible.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The ECG wave between the two green lines is the present one. Click on to select another QRS wave. Select the present lead in the pull-down list . You can choose the lead group, the gain and the speed of the waves displayed in the bottom part of the HF ECG analysis screen. is the statistic data of beadings, notches and slurs. You can change a value by double-clicking on the value.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-20 HF ECG Report 6.6.5 Analyzing HRV Click on HRV to display the HRV ECG analysis screen. The HRV ECG analysis screen includes two tabs: Auto diagnosis result and Waveform. NOTE: 1. The HRV sampling time can be set in the Sample Setting window. 2. The HRV analysis lead can be selected in the Sample Setting window.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.5.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG LF Low Frequency HF High Frequency LF/HF Ratio of low frequency to high frequency LF (norm) Standard LF power HF (norm) Standard HF power H- Doctor Diagnosis Field 1. Enter your own opinions in the Auto diagnosis textbox, and then click on the Save button. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG HRV waveform is displayed in the Waveform window (Figure 6-22). 1. You can drag the mouse in the window to choose the wave field to be printed. Then click on the Print button to print the selected wave field. 2. Click on Previous Page or Next Page to display the waves of the previous or next page. 6.6.5.3 Previewing HRV Click on the Preview button to open the HRV preview screen. is the toolbar on the HRV preview screen. 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-23 HRV Preview Screen 6.6.6 Analyzing HRT Click on HRT to display the HRT ECG analysis screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can select Average Histogram or Single Histogram in Mode View field. Click on Next PVC or Previous PVC to see the Next or Previous PVC wave and the relevant histogram. Click on the Preview button to open the HRT preview screen. is the toolbar on the HRT preview screen. 1. Click on Print(P) to print the report. 2. Click on Next Page/Prev Page to switch to the previous/next preview page. 3.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.7 Analyzing Vector ECG Click on the VCG button on the ECG analysis screen to display the VCG analysis screen. Vector ECG displays 3D image of ECG activity. You can choose the plane and the vector loop on the VCG analysis screen. Plane choices include Frontal (F), Horizontal (H), Sagittal(S) and ALL. Loop choices include P loop, QRS loop, T loop and ALL. 6.6.7.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-24 displays Vector ECG with the plane of ALL and the loop of ALL. A- Vector ECG of Frontal (F) B- Vector ECG of Horizontal (H) C- Vector ECG of Sagittal (S) D- Average templates of X, Y and Z leads. Double-click on this figure to display the magnified average template. You can drag these lines marked P1, P2, Q, S, T1 and T2 on the wave. With the change of the line position, the corresponding parameter values change.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 8 ‘8’ font ring 0.01 (amplitude) The amplitude at 0.01s from QRS loop 0.01 (angle) The angle at 0.01s from QRS loop 0.02 (amplitude) The amplitude at 0.02s from QRS loop 0.02 (angle) The angle at 0.02s from QRS loop 0.03 (amplitude) The amplitude at 0.03s from QRS loop 0.03 (angle) The angle at 0.03s from QRS loop 0.04 (amplitude) The amplitude at 0.04s from QRS loop 0.04 (angle) The angle at 0.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.7.2 Displaying Vector ECG with Frontal Plane and QRS Loop Figure 6-25 Vector ECG - Frontal & QRS loop The percent values of 0.00%, 94.16%, 1.35% and 4.49% in the square represent the area percentages of QRS loop in every quadrant. 20 mm/mV indicates the magnified multiple (gain). The red curve is QRS loop. You can click on the Zoom in button or the Zoom out button to change the gain of the displayed graphics.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Click on Zoom In/ Zoom Out to magnify/minify the picture. Click on Reset to restore the magnified/minified picture to original size. Click on Preview to preview the 3D graph. Click on Print to print the 3D graph. Click on Save to save the graph on the current screen. Click on Return to return to the ECG analysis screen. 6.6.7.4 Previewing Vector ECG Click on the Preview button to open the VCG preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-26 VCG Preview Screen (Plane is ALL & loop is ALL) 6.6.8 Analyzing Time Vector ECG Click on the TVCG button on the ECG analysis screen to display the TVCG analysis screen. Time Vector ECG is Vector ECG including time factor. As Figure 6-27 shows, you can observe waves of X lead, Y lead, Z lead, X-Y lead, X-Z lead and Z-Y lead. You can choose the speed and the gain of the displayed waves.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-27 Time Vector ECG Screen Click on the Preview button to display the TVCG preview screen. is the toolbar on the TVCG preview screen. 1. Click on the Zoom In button on the toolbar to magnify the preview page. 2. Click on the Zoom Out button on the toolbar to minify the preview page. 3. Click on the Close button to close the TVCG preview screen and return to the previous screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.9 Analyzing Signal Averaged ECG SAECG is also called Ventricular Late Potential (VLP). Click on SAECG to open the SAECG analysis screen. The SAECG analysis screen includes two tabs: Time domain and Frequency domain. 6.6.9.1 About the Time Domain Window Click on the Time domain tab to open the Time domain window. Figure 6-29 Time Domain Window Every parameter of SAECG is shown in the following table.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Select a filter in the Filter pull-down list . The top of the Time domain window is the window of standard superimposed QRS waveform. The left green line is the starting point of standard unfiltered QRS waveform, and the right green line is the end point. You can respectively drag the two green lines to change QRS duration. The bottom of the Time domain window is the window of standard superimposed QRS waveform after filtering.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.9.3 Previewing Signal Averaged ECG Click on the Preview button to display the SAECG preview screen. is the toolbar on the SAECG preview screen. 1. Click on Print(P) to print the report. 2. Click on Next Page/Prev Page to switch to the previous/next preview page. 3. Click on Two Page to preview two pages on one screen simultaneously. 4. Click on Zoom In/ Zoom Out to magnify/minify the preview page. 5.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.10 Printing ECG Reports 1. Choose Start > Printers and Faxes, and then right-click on the icon of the printer used, and select Set as Default Printer. Then close the Printers and Faxes window. 2. Click on the Print button on the analysis screen to print an ECG report. 3. Or, click on the Print button on the preview screen to print an ECG report. 6.6.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG If you select Send, the sampled data will be sent by OUTLOOK EXPRESS (Windows XP) or Window Live Mail (Windows 7/Vista) when it is saved to the designated directory. During the saving and sending course, the system will give the hint information. NOTE: In Windows 7/Vista, only if Window Live Mail is installed, can the report be sent by email. 6.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Chapter 7 Operation Instructions for Exercise ECG The exercise ECG function is optional. It is available only if you purchased this function. 7.1 Viewing Lead Placement Information 1. Click on the Lead Placement button on the main screen to display the Lead Placement window. 2. Click on Exercise ECG lead system to view the lead placement information.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2 Selecting a Patient Record to Start a New Test 1. You can select a patient record from the database to start a new test. The operation steps are the same as those of resting ECG. For details, refer to Section 6.2, “Selecting a Patient Record to Start a New Test”. 2. Select Exercise ECG in the Patient Information window. 3. Click on OK in the Patient Information window to open the Exercise ECG Setting window.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2.1 Setting Target HR The system applies the following formulas to calculate the target heart rate. 220 and 85 are default values, and you can modify them in different situations in the Exercise ECG Setting window. 7.2.2 Setting Normal BP Set the normal BP range in the Exercise ECG Setting window. When the patient’s BP exceeds the normal BP range, the system will consider it as abnormal status. 7.2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2.5 Setting BP Sampling Mode Select a BP sampling mode from the BP sampling mode list. 7.2.6 Setting BP Triggering Mode You can set the BP triggering mode to Square wave or QRS in the Exercise ECG Setting window. 7.2.7 Setting Auto Printing Select Auto print to print ECG reports automatically. 7.3 Entering New Patient Information If the patient is a new one, 1.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 2. Select Exercise ECG in the Patient information window. 3. Click on OK in the Patient information window to open the Exercise ECG Setting window. After setting the parameters, click on the OK button to open the pre-sampling screen. For details, please refer to Section 7.2, “Selecting a Patient Record to Start a New Test”. 7.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 1. Select a display mode from the display mode pull-down list 2. Select a gain from the gain pull-down list 3. Select a speed from the speed pull-down list 4. Select a lowpass filter from the lowpass filter pull-down list . . . . 7.5 Pretest Phase Click on Print to print the wave of 10 seconds before you click on Print.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.5.1 Viewing the Heart Rate and the Blood Pressure 1. View the heart rate in the heart rate field . The middle number is the current heart rate, and the right number is the target heart rate. The symbol shows the percentage of the current heart rate to the target heart rate, and the blue part indicates the current heart rate.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG NOTE: If the systolic pressure or the diastolic pressure exceeds the normal BP range, the hint The systolic/diastolic BP has exceeded the normal range! will appear on the sampling screen and the background color of the BP field will change from white to yellow. 7.5.2 Viewing Other Information 1. If a treadmill is used, view other information such as the stage time, exercise time, total time, PVC/min, speed/slope, and Max ST/Min ST. 2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Click on the Print button to print the current waveform. Click on Exit to return to the ECG sampling screen. 2. Click on the Comment button to display the Comment dialog box. Enter the comment in the Comment dialog box. Click on the OK button, and the comment will be displayed on the ECG waves on the analysis screen. 3. Click on the Template button, and the average waves of 12 leads and the calibration lines will be displayed.
SE-1010 PC ECG User Manual 4. Operation Instructions for Exercise ECG Click on the ST Trend button to display the ST trend. Right-click on the ST trend field to display the lead group menu, and then you can select other leads. 7.5.4 Printing the Pretest Report 1. The pretest report will be printed automatically 14 seconds after the beginning of the pretest phase. 2. Or, you can click on the Print button to print the pretest report. 7.6 Exercise Phase 1.
SE-1010 PC ECG User Manual 7. Operation Instructions for Exercise ECG Click on the Keep button and the system enters the Keep state. The hint Keep will be displayed in the speed/power field. The Next button becomes unavailable. In this state, the system will not follow the previous settings to change the speed and grade of the treadmill, but keep the current speed and grade until this button is pressed again. 7.7 Recovery Phase 1.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9 About Analysis Screen 7.9.1 About Summary Screen 1. Click on Summary to open the Summary screen. Figure 7-2 Summary Screen Click on the Measure button, click on one point on the wave, and then drag the mouse to another point. The distance, amplitude difference and heart rate between the two points will be displayed. 2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 5. Enter diagnosis results in the Diagnosis textbox, and then click on Save. 6. Or, double-click on the necessary results required to be added in the Glossary textbox, the selected results will be displayed in the Diagnosis textbox, and then click on the Save button. 7. Click on the Preview button to display the Print Set window. Select Summary Report or Current Wave. Click on the OK button to open the preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Figure 7-4 Current Wave Report NOTE: The diagnosis result is displayed in the diagnosis field of the summary report. 7.9.2 About ST Analysis Screen 1. Click on the ST analysis button to display the ST analysis screen.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 2. Click on Pretest to display 12-lead ST analysis waves of the pretest phase. 3. Click on Exercise to display 12-lead ST analysis waves of the exercise phase. 4. Click on Recovery to display 12-lead ST analysis waves of the recovery phase. 5. Click on Previous Page/Next Page to display 12-lead ST analysis waves of every 10 seconds. 6. Click on Max ST Ascend to display the Max ST ascending waves. 7.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.3 About All View Review Screen 1. Click on the ALL View button to display the ALL View screen. Figure 7-6 All View Screen The ALL View screen displays the ECG wave of one lead throughout the whole test, and arrhythmia will be marked in red signs. 2. Select a lead from the lead pull-down list 3. Select a gain from the gain pull-down list to view the ECG wave of the lead. .
SE-1010 PC ECG User Manual 4. Operation Instructions for Exercise ECG Click on one point on the wave and drag the mouse to another point, and then the selected range will be marked with two red lines. Then click on the Zoom in button to display the amplified ECG segment. Drag the bottom scroll bar to view the whole amplified ECG waves. 5. The ALL View screen displays ECG waves of 450 seconds (50 seconds in one line and at least 9 lines in one page).
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.4 About Trend Screen Click on the Trend button to display the STj Trend screen. Figure 7-7 STj Trend Screen You can observe 12-lead trend of the exercise test on the STj Trend screen. The horizontal coordinate indicates test time, but the unit of the vertical coordinate is different on different trend screens. Click on Zoom In or Zoom Out button to adjust the value on the vertical coordinate.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.5 About ECG Strip Screen Click on the ECG Strip button to display the ECG Strip screen. Figure 7-8 ECG Strip Screen 1. Click on Pretest to display 12-lead strip waves of the pretest phase. 2. Click on Exercise to display12-lead strip waves of the exercise phase. 3. Click on Recovery to display12-lead strip waves of the recovery phase. 4. Click on Previous Page/Next Page to display strip waves of every 30 seconds. 7.9.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.7 Printing ECG Reports You can print an ECG report by clicking on the Print button. For details, please refer to Section 6.6.10, “Printing ECG reports”. 7.9.8 Saving ECG Reports You can save an ECG report by clicking on the Report Save button. For details, please refer to Section 6.6.11, “Saving ECG reports”. 7.9.9 Exiting the Analysis Screen Click on the Exit button on the analysis screen to return to the previous screen.
SE-1010 PC ECG User Manual Processing Patient Records Chapter 8 Processing Patient Records Click on the Data Manager button on the main screen (Figure 6-1) to open the Data Manager screen (Figure 8-1). Figure 8-1 Data Manager Screen Click on a patient record in the patient information list, and then all the examination records of the patient will be displayed in the examination record list.
SE-1010 PC ECG User Manual Processing Patient Records 2. Or, select a search item in the pull-down list , enter the corresponding information in the right textbox, and then click on the Search button. All the patient records which meet the conditions will be displayed in the patient information list. 3. Or, click on Advanced Search to display the Search Condition window, and then enter the search conditions.
SE-1010 PC ECG User Manual 4. Processing Patient Records Or, you can click on Physician or Examination , and then choose the doctor name or examination types, all the patient records which meet the conditions will be displayed in the patient information list. 5. Select The patients with no examination, and then the patient records which are registered but not examined, will appear in the patient information list. 8.
SE-1010 PC ECG User Manual Processing Patient Records 8.3 Deleting Records NOTE: The deletion of records is permanent, and you can’t restore the records deleted. Please use this operation cautiously. 8.3.1 Deleting Patient Records Click on a patient record in the patient information list on the Data Manager screen, and then click on the Delete button to delete the patient record from the patient information list. At the same time, all the examination records of the patient will be deleted.
SE-1010 PC ECG User Manual Processing Patient Records 8.5 Merging Examination Records Click on one or more examination records in the examination record list on the Data Manager screen, and then click on the Merge/Assign button to display the Patient Information window. Input a patient ID and click on the OK button to assign the examination record selected to this patient. 8.
SE-1010 PC ECG User Manual Processing Patient Records When Template is selected, the templates of the two examination records will be displayed in the window. You can press Print button to print the current window. Click on the Ruler button on the Compare window. Click on one point on the wave, and then drag the mouse to another point. The distance, amplitude difference and heart rate between the two points will be displayed.
SE-1010 PC ECG User Manual Processing Patient Records 8.7 Importing ECG Data into the Data Manager Screen Click on the Import button on the Data Manager screen (Figure 8-1) to open the following window. Select the data to be imported and click on the Select button to import the data into the Data Manager screen.
SE-1010 PC ECG User Manual Processing Patient Records If the data to be imported exists on the Data Manager screen, the following hint will pop up. If you press the Yes button, the imported record will replace the file with the same name. If you press the No button, the system will hint you a failure operation occurs. NOTE: Only ECG data in DAT format can be imported.
SE-1010 PC ECG User Manual Processing Patient Records 8.8 Exporting ECG Data from the Data Manager Screen Select examination records and click on the Export button on the Data Manager screen (Figure 8-1) to open the following window. Assign the file name, saving path and export file format (SCP, FDA-XML, DICOM, dat, pdf), and then click on the OK button to export the data into the selected path. At the same time, the patient information of these records will be exported.
SE-1010 PC ECG User Manual Processing Patient Records 8.9 Viewing an Examination Record Click on a patient record in the patient information list, and then all the records of the patient will be displayed in the examination record list. Select All records and all the examination records will be displayed in the examination record list. Double-click on an examination record in the examination record list on the Data Manager screen (Figure 8-1).
SE-1010 PC ECG User Manual Configuring the System Chapter 9 Configuring the System Click on the System Setting button on the main screen (Figure 6-1) to open the System Setting window. There are eight tabs in the System Setting window: Basic Information, Sample Setting, Device, Print Setting, Output File, Data Maintenance, GDT and Others. After you modify some information in the System Setting window, 1. Click on the OK button to save these modifications and exit. 2.
SE-1010 PC ECG User Manual Configuring the System 9.1.1 Setting Basic Information Enter information in the Hospital Name, User-Defined 1 or User-Defined 2 textbox in the Basic Information window (Figure 9-1). When you fill in the User-Defined 1/2 textbox, the corresponding items in the Patient Information window will change into what is filled.
SE-1010 PC ECG User Manual Configuring the System NOTE: Click on the New Patient button on the main screen to open the Patient Information window as the above figure shows. Select Memory in Basic Information window, the content of User-Defined 1 in the Patient Information window will be saved.
SE-1010 PC ECG User Manual Configuring the System 9.2 Sample Setup Figure 9-2 Sample Setup Window 9.2.1 Setting Filter Set filters in the Sample Setting window (Figure 9-2).
SE-1010 PC ECG User Manual Configuring the System DFT Filter DFT filter greatly reduces the baseline fluctuations without affecting ECG signals. There are two options: Weak and Strong. NOTE: If DFT filter is set to Strong, the ECG data displayed on the screen is 0.85 seconds later than the real-time ECG data; if DFT filter is set to Weak, the ECG data displayed on the screen is 1.8 seconds later than the real-time ECG data. EMG Filter EMG filter suppresses the disturbance caused by strong muscle tremor.
SE-1010 PC ECG User Manual Configuring the System ♦ HR Calculation You can set the HR calculation to one of the 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. ♦ HR Analysis Lead You can set the HR Analysis Lead to one of the 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. 9.2.
SE-1010 PC ECG User Manual Configuring the System 9.3 Device Setup Figure 9-3 Device Setup Window 9.3.1 Setting Sampling Device Select a sample device from the Sampling Device pull-down list in the Device window (Figure 9-3).
SE-1010 PC ECG User Manual Configuring the System When DX12 wireless sampling box is selected, you can click on the Address button to view the receiver address. You can match the transmitter and receiver based on the address from the Sampling Device pull-down list. 9.3.2 Setting Device Type/Mode Select a device type/mode from the Device Type/Device pull-down list . 9.3.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.1 Creating a New Protocol Click on the New button in the Device window (Figure 9-3) to enter the Edit the protocol window. 1. Enter the protocol name in the Protocol Name textbox. 2. Right-click on the grid to display the menu , click on Add Stage, and enter information in the grid. 3. Set the auto printing time of 12-lead ECG reports in the 12-lead ECG Report (First) and 12-lead ECG Report (Repeat) columns. 4.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.2 Modifying a Protocol 1. Select a protocol from the Protocol pull-down list, and then click on the Modify button to display the Edit Protocol window. 2. Double-click on an option of the phase row to be modified in the protocol list, and the pull-down list the corresponding grid. will be displayed. Select a phase and it will be displayed in 3.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.3 Deleting a Protocol Select a protocol from the Protocol pull-down list, and then click on the Delete button to delete the protocol. NOTE: You can’t delete default protocols Bruce and IsoPower. 9.3.4.4 Restoring the default protocol Click on the Restore the default button to resume the default protocol. 9.3.5 Advanced Setup Click on Advanced Setting, input the correct password in the pop-up textbox, and then click on the OK button. 9.3.5.
SE-1010 PC ECG User Manual Configuring the System Only if Enable System Integration is selected, can Device ID and Output Path be set. Device ID and Output Path should be set to the same values as those of Smart ECG Net. Otherwise, data of this system can not be uploaded to Smart ECG Net system normally.
SE-1010 PC ECG User Manual Configuring the System 3. Click on Device Manager on the Computer Management screen, and then click on Human Interface Devices to view port information. 4. In the Barcode Setting window, set Device to the port you view on the Computer Management screen, and then click on the OK button to confirm. 5. Restart the PC ECG software.
SE-1010 PC ECG User Manual Configuring the System 9.4 Print Setup Figure 9-4 Print Setting Window 9.4.1 Choosing Patient Information to be Printed The default items of the patient information are Department, Room No and Physician. You can also select the additional information, such as FirstName/LastName, Date of Birth, height, weight, BP, race, medication, Pacemaker, Technician and Ref-Physician.
SE-1010 PC ECG User Manual Configuring the System If you select First Name/Last Name, the Patient Name textbox in the Patient Information window will change into the First Name and Last Name textboxes. If you select D.O.B, the D.O.B textbox appears in the Patient Information window, and the Age textbox becomes unavailable. 9.4.2 Choosing Diagnosis Information to be Printed The diagnosis information is displayed on the preview screen and in the report printed out.
SE-1010 PC ECG User Manual Configuring the System 9.4.3 Setting Rhythm Lead The rhythm lead can be one of 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. When the printing mode is set to 3×4+1 or 6×2+1, the rhythm lead selected in the Rhythm1 list box will be printed out. When the printing mode is set to 3×4+3, 3 rhythm leads selected in the Rhythm1, Rhythm2 and Rhythm3 list boxes will be printed out. 9.4.4 Defining Printing Format 1.
SE-1010 PC ECG User Manual Configuring the System 9.5 Output File Setup Figure 9-5 Output File Setup Window 9.5.1 File Naming The default file name is Exam Time-ID, and a “-” exists between every two fields. You can also modify the file name manually, such as Name-ID-Exam Time. Each field includes ID, name, exam time, age and sex items, you can randomly use these items to combine a file name, but you should select at least one item as a file name.
SE-1010 PC ECG User Manual Configuring the System 9.5.2 Setting PDF/JPG Select Output PDF when making diagnosis, the system will automatically output files in PDF format when making diagnoses. Select Output JPG when making diagnosis, the system will automatically output files in JPG format when making diagnoses. 9.5.3 Setting SCP Select Data Compression, the system will automatically output files in SCP format of data compression.
SE-1010 PC ECG User Manual Configuring the System 9.5.5 Setting DICOM Select Output when sampling finishes, the system will automatically output files in DICOM format when sampling finishes. Deselect Output when sampling finishes, the system will not output files in DICOM format when sampling finishes. Select Output when making diagnosis, the system will automatically output files in DICOM format when making diagnoses.
SE-1010 PC ECG User Manual Configuring the System 9.6.1 Database Rebuild You can rebuild the database to avoid losing data because of damaged data files. Click on the Rebuild Database button in the Data Maintenance window, and then you can select the path to rebuild database. You cannot enter the path manually. After rebuilding the path successfully, hint information will be displayed. 9.6.2 Database Backup You can also make a backup for the data files to avoid your data files being damaged or missing.
SE-1010 PC ECG User Manual Configuring the System Click on the OK button in the Backup Setting window, the system will begin backup. You can check the backup status according to the progress bar or you can stop backup at any times. The system will back up the data files for the latest time if you do not change the path.
SE-1010 PC ECG User Manual Configuring the System You can also set the reminded backup period to remind you back up the data files at any time. The default system reminded period is None, you can also select 7 days, 14 days, 30 days or you can define it manually. 9.
SE-1010 PC ECG User Manual Configuring the System Select GDT to enable GDT features. Click on the Browse button, and then appoint the path to exchange files with EDP. Fill in the Input file name textbox to set command file name assigned by EDP to the PC ECG system. Fill in the Output file name textbox to set data file name that is used by the PC ECG system to transmit the result data to EDP. Enter ECG ID to set GDT ID assigned to the system which will be entered in field 8315 or 8316 in the GDT protocol.
SE-1010 PC ECG User Manual Configuring the System 9.8.1 Setting Unit and Color Set the height unit to cm, inch or mm. Set the weight unit to Kg, g or Pound. Set the BP unit to kPa or mmHg. Set the speed unit to Km/h or mph. Set the grade unit to % or degree. Set the time mode to 24Hour or 12Hour. Set the date mode to MM-DD-YYYY, DD-MM-YYYY OR YYYY-MM-DD. Set the color of the background, waves, grid (5mm), grid (1mm), mark and text.
SE-1010 PC ECG User Manual Configuring the System 9.9 Modifying the Glossary Click on Edit Diagnosis Template from the Tool (F) pull-down list on the main screen (Figure 6-1), and then the Edit Diagnosis Template window appears. 1. Adding an item Enter a diagnosis item, such as aa in the textbox, and then click on the Add button. The added item will be displayed in the Edit Diagnosis Template window.
SE-1010 PC ECG User Manual 2. Configuring the System Adding a subitem Click on the item you wanted to add a subitem, enter a diagnosis subitem, such as bb in the textbox, and then click on the Add button. The added subitem will be displayed under aa. 3. Deleting an item Click on the item you wanted to delete from the Edit Diagnosis Template window, and then click on the Delete button to delete this item. 4. Save the settings Click on the Save button to save these modifications.
SE-1010 PC ECG User Manual Hint Information Chapter 10 Hint Information Hint information and the corresponding causes provided by the system are listed as follows. Table 10-1 Hint Information and Causes Hint Information Causes Lead off: X Electrodes fall off the patient or the patient cable falls off the ECG sampling box. It is pre-sampling now, please click on During the pre-sampling course ‘Start’ to begin recording.
SE-1010 PC ECG User Manual Hint Information Sampling Device is in sleep mode, please DX12 transmitter is in sleep mode. press “Power” to wake it up. Overload The direct current offset voltage on an electrode is too high. Sorry, Can not Connect to the Database! MSDE 2000 or SQL Server 2005 Express is not started up. Fail to create database! The system fails to create database.
SE-1010 PC ECG User Manual Cleaning, Care and Maintenance Chapter 11 Cleaning, Care and Maintenance CAUTION Turn off the system power and drag the power cable out from the socket before cleaning or disinfection. 11.1 Cleaning and Maintaining the Treadmill Daily Cleaning and Maintenance 1. Wipe the treadmill with a clean soft cloth to remove dust, moisture and sweat stain. 2. Wipe the handrail of the treadmill with a clean soft cloth damped in non-caustic neutral detergent. 3.
SE-1010 PC ECG User Manual Cleaning, Care and Maintenance ♦ Do not drag or twist the patient cable with excessive stress while using it. Hold the connector plugs instead of the cable when connecting or disconnecting the patient cable. ♦ Align the patient cable to avoid twisting, knotting or crooking at a closed angle while using it. ♦ Store the lead wires in a big wheel. ♦ Once damage or aging of the patient cable is found, replace it with a new one immediately.
SE-1010 PC ECG User Manual Accessories Chapter 12 Accessories WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Table 12-1 Standard Accessory List for Wired System Accessory Part Number DP12 ECG Sampling Box 02.01.210039 Sentinel / USB 01.18.047225 Resting ECG External USB Cable 01.13.036134 DP12 Patient Cable / European Standard 01.57.
SE-1010 PC ECG User Manual Accessories Table 12-3 Optional Accessory List Accessory Part Number Sentinel / USB 01.18.047229 Patient Cable for Exercise ECG (only for Wired ECG System) 01.57.109850 Resting ECG External USB Cable 01.13.036134 Disposable Frosting Film for Skin Preparation 01.57.107418 MSB1212 Disposable Electrode 01.57.040171 Exercise ECG External USB Cable 01.13.036135 Computer Lenovo, Qitian M6900 Snap/Banana Socket Adapter (only for Wired ECG System) 11.18.052208 01.13.
SE-1010 PC ECG User Manual Warranty & Service Chapter 13 Warranty & Service 13.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: a) damage caused by mishandling during shipping. b) subsequent damage caused by improper use or maintenance. c) damage caused by alteration or repair by anyone not authorized by EDAN.
SE-1010 PC ECG User Manual Recommended Optional Accessories Chapter 14 Recommended Optional Accessories Treadmill: Model: TM-400 Manufacturer: EDAN INSTRUMENTS, INC. China CE marking Model: Valiant Manufacturer: Lode B.V. The Nettherlands CE marking Model: h/p/cosmos (all medical models) with coscom interface Manufacturer: Full Vision Inc. USA CE marking Model: mercury med 4.0, mercury 4.
SE-1010 PC ECG User Manual Recommended Optional Accessories STRESS BP: Model: Tango+ Manufacturer: SunTech Medical Inc. USA CE Certificate and FDA 510(k) clearance Isolating Transformer: Model: ES710 Manufacturer: BenDer Inc. Deutschland Electrical Outlet: Power Consumption: no less than 4500VA Special use for medical equipment Printer: Model: HP2010, HP2035 Manufacturer: HP Company, USA Model: CANON iP1980 Manufacturer: CANON Company, Japan WARNING 1.
SE-1010 PC ECG User Manual Technical Specifications Appendix 1 Technical Specifications A1.
SE-1010 PC ECG User Manual Technical Specifications A1.3 Physical Specifications DP12 ECG sampling box: 148 mm (L) ×100 mm (W) × 40 mm (H) (5.8in×3.9in×1.6in) Dimensions DX12 transmitter: 63mm(L)×107mm(W) ×23mm(H) (2.5in×4.2in×0.9in) DX12 receiver: 155mm(L)×100mm(W)×30mm(H) (6.1in×3.9in×1.2in) DP12 ECG sampling box: Approx. 210g Weight DX12 transmitter: Approx. 113g (not including battery) DX12 receiver: Approx. 173g A1.
SE-1010 PC ECG User Manual Technical Specifications Paper Speed: 5 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s(±3%) Record message: Date, Time, Printing Speed, Filter, Symbol, Heart Rate, Patient ID, Sex, Age, Lead Mark, Lead Wave, Average Template Wave or Rhythm Lead Wave, Measurement Result and Interpretation Information Result (option) etc.
SE-1010 PC ECG User Manual Technical Specifications Work Frequency Filter DFT Filter: weak/strong LOWPASS Filter: 25 Hz / 35 Hz / 45 Hz / 75 Hz / 100 Hz / 150 Hz CMRR DP12 ECG sampling box≥110 dB DX12 transmitter≥100 dB Pacemaker Detection Amplitude DP12 ECG sampling box: ±2 to ±700 mV DX12 transmitter: ±2 to ±500 mV Width 0.1 to 2.0 ms Sampling Frequency 10,000 /sec/channel NOTE: Test the accuracy of input signal reproduction according to the methods described in clause 4.2.7.
SE-1010 PC ECG User Manual EMC Information Appendix 2 EMC Information Guidance and manufacture’s declaration - electromagnetic emissionsfor all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic emission SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
SE-1010 PC ECG User Manual EMC Information Guidance and manufacture’s declaration - electromagnetic immunity for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic immunity SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
SE-1010 PC ECG User Manual EMC Information 70% UT SE-1010 PC ECG be powered from an uninterruptible power supply or a battery. (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 sec NOTE UT is the a.c. mains voltage prior to application of the test level.
SE-1010 PC ECG User Manual EMC Information Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations.
SE-1010 PC ECG User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and SE-1010 PC ECG SE-1010 PC ECG is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-1010 PC ECG User Manual Abbreviation Appendix 3 Abbreviation Abbreviation Statement ECG Electrocardiograph/Electrocardiogram VCG Vector ECG TVCG Time Vector ECG QTD QT Dispersion FCG Frequency ECG HRV Heart Rate Variability HFECG High Frequency ECG HRT Heart Rate Turbulence SAECG Signal Averaged ECG TO Turbulence Onset TS Turbulence Slope VLP Ventricular Late Potential HR Heart Rate P Dur P-wave Duration PR Dur P-R Interval QRS Dur QRS Complexes Duration QT/QTc Q-T In
SE-1010 PC ECG User Manual Abbreviation HF (norm) Standard HF Power aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm R Right RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus - 163 -