Product Manual

M3A Vital Signs Monitor User Manual Specifications
- 88 -
A1.2.6 Review
Trend List Review 100 hours, 30 seconds Resolution
Measurement Review 12, 000 groups measuring data
A1.2.7 NIBP (Optional)
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring interval in AUTO mode 1/2/3/4/5/10/15/30/60/90/120/240/480 min
Continuous 5 min, interval is 5s
Measuring type Systolic Pressure, Diastolic Pressure, Mean Pressure
Measuring range
SYS 40 mmHg ~ 270 mmHg
DIA 10 mmHg ~ 215 mmHg
ADU mode
MAP 20 mmHg ~ 235 mmHg
SYS 40 mmHg ~ 200 mmHg
DIA 10 mmHg ~ 150 mmHg
PED mode
MAP 20 mmHg ~ 165 mmHg
SYS 40 mmHg ~ 135 mmHg
DIA 10 mmHg ~ 100 mmHg
NEO mode
MAP 20 mmHg ~ 110 mmHg
Alarm type SYS, DIA, MAP
Cuff Pressure measuring range 0 mmHg ~ 300 mmHg
Pressure resolution 1 mmHg
Maximum mean error
±5 mmHg
Maximum standard deviation 8 mmHg
Maximum measuring time of single
measurement
ADU/PED 120s
NEO 90s
Typical measuring period 30s ~ 45s (depend on HR/motion disturbance)
Overpressure protection Dual overpressure protection