Product Manual
M3A Vital Signs Monitor User Manual Specifications
- 85 -
Appendix SpecificationⅠ s
A1.1 Classification
Anti-electroshock type Class equipment and internalⅠ ly powered equipment
EMC type Group I Class A
Anti-electroshock degree
SpO
2
, NIBP: BF Defibrillation type;
TEMP: CF type (T2 module)
BF type (TH module)
Ingress Protection
IPX1 (No protection against ingress of water if
configured with TEMP module)
Working system Continuous running equipment (no more than 7 days)
Compliant with Safety Standards IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+
A2, IEC/EN 60601-1-2:2001+A1, IEC 60601-1-8,
ISO 9919, EN 1060-1, EN 1060-3, EN 1060-4,
ANSI/AAMI SP10, IEC/EN 60601-2-30,
IEC60601-2-49, EN 12470-4, EN 12470-5
A1.2 Specifications
A1.2.1 Size and Weight
Size 200.8 mm (L)×241 mm (H)×189 mm (D)
Weight 2.4 kg (without battery)
A1.2.2 Environment
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
Temperature
+5 °C ~ +40 °C
Working
With TEMP: +10 °C ~ +40 °C