M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 80 -
11.2.6 Maintenance and Cleaning
Calibration Mode
To switch to calibration mode, follow the steps below:
a Press the ON/MEM button to turn the thermometer on. The display of the thermometer shows
symbols and functions.
b Keep pressing the ON/MEM button for five seconds and you will see the “OFF” symbol on
the display. Do not release the button until you see a dot onscreen.
c The thermometer is now in the Calibration Mode and the display is flashing and showing the
“CAL” symbol.
NOTE:
It is suggested that a re-test is performed for the device on accuracy after three years.
Please send the complete device to the dealers or nearest service address. However, if
this device is used according to the operation instructions, periodic re-calibration is not
required.
Cleaning
The probe is the most delicate part of the thermometer. Use it with care when cleaning the lens to
avoid damage.
If the device is accidentally used without a probe cover, clean the probe as follows:
1. After the measurement, use the cotton swab moistened with alcohol (70% concentration) to
clean the lens (on the inside of the probe).
2. Allow the probe to fully dry for at least one minute.