11.2 TEMP Monitoring with TH Module
11.2.1 Introduction
M3A with the TH module (Intrared Ear Temperature Module) takes a temperature in the ear.
Diagram of the Intrared Ear Thermometer
WARNING
1 Keep the probe covers away from children.
2 Do not resue the disposable probe covers.
3 Only use the disposable probe covers supplied or recommended by EDAN. Use of
other manufacturer’s probe covers, reuse of disposable probe covers or absence of
probe covers may produce temperature measurement errors and/or inaccuraies.
4 The intrared ear thermometer is not intended for neonatal patients.
CAUTION
1 Keep the probe window clean, dry, and undamaged at times to ensure accuate
measurements. To protect the probe window, always keep the thermometer in the
storage cover while transporting or when not in use.
2 Proper installation of the probe cover ensures accurate measurements.
3 Do not autoclave.
4 The thermometer is not waterproof. Do not immerse or drip fluids on it. Should this
occur, dry the thermometer with warm air. Check for proper operation and accuracy.