M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 73 -
Message Cause Alarm Level What to do
TEMP
MODULE ERR
TEMP module self check
failure
High
Put the sensor into the sensor
bracket, take it out and measure
again. If the problem persists,
stop using measuring function
of TEMP module, notify
biomedical engineer or
Manufacturer’s service staff.
Prompt:
Message Cause What to do
Ready to TEMP
predict
The monitor prompts it after taking
the sensor out of the bracket and
warm-up is over.
Put the sensor to the measuring
position and start measuring.
TEMP Predict
complete
After the Predict measuring is over,
the data and message display on the
interface.
Enter monitoring state after the
Predict state. After monitoring for
10 min, it returns to waiting state.
11.1.5 Care and Cleaning
WARNING
Before cleaning the monitor or the probe, make sure that the equipment is switched off
and disconnected from the power line.
Reusable TEMP Probes
1 The TEMP probe should not be heated above +100 (+212 ). It should only be briefly
exposed to temperatures between +80 ~ +100 (+176 ~ +212).
2 The probe must not be sterilized in steam.
3 Only detergents containing no alcohol can be used for disinfection.
4 All the sensors should be used with a protective rubber cover.
5 To clean the probe, hold the tip with one hand and rub the probe down from the connector
with the other hand using a moist lint-free cloth.
NOTE:
1 Wash the probe with clean water after disinfecting and sterilizing to remove any
remaining solution. The probe can only be reused after being dried thoroughly.
2 Do not disinfect the probe by means of boiled water.
3 The product has not been disinfected at the factory.