M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 71 -
Figure 11-3 Alarm Setup Menu
TEMP: set it to ON to enable prompt message during the TEMP alarm, while set to OFF to
disable the alarm function, and display the symbol besides TEMP numeric.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
Set High for the higher alarm limit, and set Low for the lower alarm limit.
The range for higher alarm limit and lower alarm limit is as follows:
Patient Type Measure position High Low Step
ADU Oral/Axillary/Rectal
+42 (+107.6 ) +35.5 (+95.9 ) + 0.1
PED Oral/Axillary/Rectal
+42 (+107.6) +35.5 (+95.9 ) + 0.1
11.1.4 Alarm Message
Tables below describe the possible physiological alarms and technical alarms occurring during
TEMP measurement.
Physiological alarms:
Message Cause Alarm Level
TEMP TOO HIGH Measuring value of TEMP is above upper alarm limit. Medium
TEMP TOO LOW Measuring value of TEMP is below lower alarm limit. Medium