M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 67 -
To replace the rubber bag in the cuff, first place the bag on top of the cuff so that the rubber tubes
line up with the large opening on the long side of the cuff. Now roll the bag lengthwise and insert
it into the opening on the long side of the cuff. Hold the tubes and the cuff and shake the
complete cuff until the bag is in position. Thread the rubber tubes from inside the cuff, and out
through the small hole under the internal flap.
Disposable Blood Pressure Cuffs
Disposable cuffs are intended for one-patient use only. Do not use the same cuff on any other
patient. Do not sterilize or use autoclave on disposable cuffs. Disposable cuffs can be cleaned
using soap solution to prevent infection.
NOTE:
For protecting environment, the disposable blood pressure cuffs must be recycled or
disposed of properly.