M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 65 -
Message Cause Alarm Level What to do
MEASURE
TIMEOUT
Measuring time has
exceeded 120s (adult) or
90s (neonatal).
Low
Measure again or use other
measuring methods.
INIT
PRESSURE
TOO HIGH
The initial pressure is too
high during measuring
Low
Measure again, if failure
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
manufacturer’s service staff.
PRESSURE
RANGE
EXCEED
The measured pressure
exceeds the limit.
Low Measure by other method.
Prompt message:
Message Cause
Press NIBP START You can start NIBP measuring of continual mode.
Manual measuring... During manual measuring mode.
Automatic measuring... During automatic measuring mode.
Continual measuring... During continual measuring mode.
Measurement over Measurement over
Calibrating... During calibrating
Calibration over Calibration over
Leakage testing... During leakage test
Leakage test over Leakage test over
NIBP Resetting... NIBP module is resetting
NIBP Alarm Off The alarm of NIBP is turned off.