Product Manual

M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 64 -
Technical alarms: (display in the area below the NIBP value):
Message Cause Alarm Level What to do
NIBP COMM
STOP
NIBP module failure or
communication failure.
High
Stop using measuring function
of NIBP module, notify
biomedical engineer or
manufacturer’s service staff.
NIBP
ILLEGAL
RESET
The hardware pressure is
too high
High
NIBP
MODULE ERR
The NIBP module has
failure.
High
Measure again, if failure
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
manufacturer’s service staff.
LOOSE CUFF
Cuff is no properly
wrapped or no cuff exists.
Low Properly wrap the cuff.
AIR LEAK
Cuff, hose or connector is
damaged.
Low
AIR
PREESURE
ERR
The airway of NIBP has
failure.
Low
Check and replace the leaking
parts, if required, notify
biomedical engineer or
manufacturer’s service staff.
NIBP SIGNAL
TOO WEAK
Cuff is too loose or patient
pulse is too weak.
Low
Use other method to measure
blood pressure.
NIBP NOISE
SIGNAL
Because of arm motion,
signal noise is too large or
pulse rate is not regular.
Low
Make sure that the patient
under monitoring is
motionless.
OVER
PRESSURE
Pressure has exceeded the
specified upper safety
limit.
Low
Measure again, if failure
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
manufacturer’s service staff.
NIBP SIGNAL
SATURATED
Excessive motion. Low Stop the patient from moving.
CUFF TYPE
ERR
Cuff type does not comply
with the patient type.
Low Select appropriate cuff type