8.5 Replacement of Fuse
Unscrew the fuse cap anticlockwise, replace the fuse (protector tube) and screw down the fuse
cap clockwise. Fuse size: Ф5×20, Rated value: T 1.6 AL /250 V.
NOTE:
Switch off the power of the monitor before examining the fuse.
8.6 Cleaning Battery and Battery Compartment Cover
Use only non-caustic detergents such as soap and warm water (+40℃/+104℉ maximum) to
clean the battery. Do not use strong solvent to clean battery, and do not dip the battery in liquid.