Product Manual

M3A Vital Signs Monitor User Manual System Menu
- 32 -
Figure 4-17 SpO
2
Setup
Sensitivity
The SpO
2
reading is the average of data collected within a specific time. You can set
Sensitivity to Low, Medium or High via the menu. The higher the sensitivity is, the quicker
the pulse oximeter responds to the changes in the patient’s oxygen saturation level. Contrarily,
the lower the sensitivity is, the slower the pulse oximeter responds to the changes in the
patient’s oxygen saturation level, but the measurement accuracy will be improved. When a
critical patient is monitored, selecting high sensitivity will help to understand the patient’s
state.
Sensor Off
You can configure the alarm level for SpO
2
Sensor Off to Low or High.
Alarm Setup
Access Alarm Setup and you can see the menu below:
Figure 4-18 Alarm Setup (for monitor with recorder)
Alarm Mute: Set the duration of silencing the audible alarm to 1 min, 2 min or 3 min.
Alarm Silence
You can set this item to ON or OFF. If the item is ON, you can turn off the audio system by
pressing the SILENCE button on the front panel for more than 2s. In this case, all sounds
including the alarm sound, key sound and sphygmic sound coming from the monitor will be
mute. If the item is OFF, the function mentioned above is unavailable.
Alarm Record (only available for the monitor outfitted with a recorder)
By configuring Alarm Record, the function of automatically outputting the alarm
information in case of any physiological alarm can be enabled or disabled. If the item is ON,
the monitor will automatically print out the alarm information once any physiological alarm
is triggered. If the item is OFF, the monitor will not automatically output the alarm
information.