M3A Vital Signs Monitor User Manual System Menu
- 29 -
User Maintenance
Input the user password 9 9 8 1 in the Enter Password box and press Confirm:
Figure 4-12 Enter the Password
User Maintenance menu will pop up, in which you can set the following items.
Figure 4-13 User Maintenance
Language: Set the displaying language.
NOTE:
The user should restart the monitor after changing the displaying language.
Nurse Call: Turn on or off the nurse call. When the parameter alarm occurs, the monitor
gives a 3s nurse call alarm prompt; if the audio alarm or the audio system is off, the monitor
can also give the nurse call alarm in abnormal condition.
The relay contact between pin7 and pin8 of RJ45 is normally open. But it is closed when an
alarm is audible.
NIBP Setup
Access NIBP Setup and you can see the menu as follows:
Figure 4-14 NIBP Setup