Product Manual

M3A Vital Signs Monitor User Manual Intended Use and Safety Guidance
- 4 -
WARNING
21 The simultaneous use of cardiac pacemaker and other patient-connected equipment
may cause safety hazard.
22 Please disinfect timely to prevent cross infection between patients.
23 This monitor is not a device for treatment purposes.
24 Only NIBP and SpO
2
applied parts of the monitor are defibrillation-proof. When a
defibrillator is applied, keep other accessories away from the patient. Otherwise, it
may result in damaging the monitor or harming the patient.
25 Do not touch the patient, bed or instrument during defibrillation.
26 During monitoring, if the power supply is off and there is no battery for standby, the
monitor will be off, and only the patient information and alarm settings can be saved.
After reconnecting the power supply, the user should turn on the monitor for
monitoring.
CAUTION
1 Federal law (U.S.) restricts this device to sale by or on the order of a physician.
2 Electromagnetic Interference – Ensure the environment in which the monitor is
installed is not subject to any sources of strong electromagnetic interference, such as
radio transmitters, mobile telephones, etc.
3 The monitor is designed for continuous operation and is “ordinary” (i.e. not drip or
splash-proof).
4 Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine, dust
area, high-temperature and humid environment.
5 Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
6 Do not sterilize the monitor, recorder or any accessories.
7 The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do not dispose them
together with house-hold garbage. At the end of their life hand the batteries over to
the applicable collection points for the recycling of waste batteries. For more detailed
information about recycling of this product or battery, please contact your local Civic
Office, or the shop where you purchased the product.
8 Remove a battery whose life cycle has expired from the monitor immediately.