- 3 -
WARNING
6 SHOCK HAZARD- the power receptacle must be a three-wire grounded outlet. A
hospital grade outlet is required. Never adapt the three-prong plug of the monitor to fit
a two-slot outlet.
7 Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN60950 for data processing
equipment and IEC/EN60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard
IEC/EN60601-1-1. Everybody who connects additional equipment to the signal input
connector or signal output connector configures a medical system, and is therefore
responsible that the system complies with the requirements of the valid version of the
system standard IEC/EN60601-1-1. If in doubt, consult our technical service
department or your local distributor.
8 Use the battery only in this monitor. Do not connect battery directly to an electric outlet
or cigarette lighter charger.
9 Do not unplug the battery when monitoring.
10 Make sure the monitor is used in the appointed range of voltage, the effect of power
supply can be ignored.
11 Do not solder the leading wire and the battery terminal directly.
12 If liquid leaking from the battery gets into your eyes, onto your skin or clothes, do not
rub your eyes. Wash them well with clean water and go to see a doctor immediately.
13 Always keep the battery away from fire.
14 Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
15 Do not use a battery with serious scar or deformation.
16 Only patient cable and other accessories supplied by EDAN can be used. Or else, the
performance and electric shock protection can not be guaranteed, and the patient
may be injured.
17 Please set the alarm according to the individual condition of patient to avoid delaying
treatment. Ensure there will be an alarm audio prompt when an alarm occurs.
18 Devices connecting with the monitor should be equipotential.
19 When the monitor and electrosurgical device are used together, the user (physician or
nurse) should guarantee the safety of patient.
20 The monitor can provide protective means to prevent the patient from being burned
when used with HF SURGICAL EQUIPMENT.