User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance
- 3 -
WARNING
10. Electrodes of dissimilar metals should not be used; otherwise, they may cause
a high polarization voltage.
11. The disposable electrodes can only be used for one time.
12. Do not touch the patient, bed, table, or device while using the ECG and a
defibrillator.
13. Do not touch accessible parts of electrical equipment and the patient
simultaneously.
14. Disposable electrodes must be used during defibrillation.
15. The use of equipment that applies high-frequency voltages to the patient
(including electrosurgical equipment and some respiration transducers) is not
supported and may produce undesired results. Disconnect the patient data
cable from the electrocardiograph, or detach the leads from the patient before
performing any procedure that uses high-frequency surgical equipment.
16. This device complies with FCC RF radiation exposure limits set forth for an
uncontrolled environment. This device must operate with a minimum distance
of 20 cm between the radiator and user body.
17. This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
a) This device may not cause harmful interference, and
b) This device must accept any interference received, including interference
that may cause undesired operation.
18. Pay close attention to the examination to avoid missing necessary ECG
waveforms.
19. SHOCK HAZARD - Don't connect non-medical electrical equipment, which has
been supplied as a part of the system, directly to the wall outlet when the
non-medical equipment is intended to be provided by a multiple portable
socket-outlet with an isolation transformer.
WARNING