SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information
- 72 -
Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM
Recommended separation distances between portable and mobile RF communications
equipment and the SE-1200 Pro&SE-1201 Pro ECG
The SE-1200 Pro&SE-1201 Pro ECG is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the SE-1200
Pro&SE-1201 Pro ECG can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
SE-1200 Pro&SE-1201 Pro ECG as recommended below, according to the maximum output power
of the communications equipment.
The maximum
output power
rating of the
transmitter
(W)
Separation distance according to the frequency of the transmitter (m)
150 kHz to 80 MHz
Pd 2.1=
80 MHz to 800 MHz
Pd 2.1=
800 MHz to 2.7 GHz
Pd 3.2=
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.