Electromagnetic Immunity
Guidance and manufacturer’s declaration - electromagnetic immunity
The SE-1200 Pro&SE-1201 Pro ECG is intended for use in the electromagnetic environment
specified below. The customer or the SE-1200 Pro&SE-1201 Pro ECG user should ensure that it is
used in such an environment.
Immunity test
IEC/EN 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC/EN
61000-4-6
3 V
rms
150 kHz to 80 MHz
6Vrms
c)
in ISM bands
between
0.15 MHz and 80 MHz
3V
rms
150 kHz to
80 MHz
6Vrms
c)
in
ISM bands
between 0.15
MHz and 80
MHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the SE-1200 Pro&SE-1201
Pro ECG, including cables, than the
recommended separation distance
calculated from the equation applicable
to the transmitter's frequency.
Recommended separation distance
Pd 2.1=
150 kHz to 80 MHz
Radiated RF
IEC/EN
61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
80 MHz to
2.7 GHz
Pd 2.1=
80 MHz to 800 MHz
Pd 3.2=
800 MHz to 2.7 GHz
See Table 1.
See Table 1.
E/6 Pd =
At RF wireless
communications equipment bands
(Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to
any part of the SE-1200 Pro&SE-1201
Pro ECG, including cables specified by
the manufacturer). Otherwise,
degradation of the performance of this
equipment could result.