User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information
- 66 -
Electromagnetic Immunity
Guidance and manufacturers declaration - electromagnetic immunity
The SE-1200 Pro&SE-1201 Pro ECG is intended for use in the electromagnetic environment
specified below. The customer or the SE-1200 Pro&SE-1201 Pro ECG user should ensure that it is
used in such an environment.
Immunity test
IEC/EN 60601
test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC/EN 61000-4-2
2, 4, 6, 8
kV contact
2, 4, 8, 15
kV air
2, 4, 6, 8 kV
contact
2, 4, 8, 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
the floor is covered with
synthetic material, the
relative humidity should be
30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
AC port: 2kV
Signal port:
1kV
AC port: 2kV
Signal port: 1kV
The main power quality
should be a typical
commercial or hospital
environment.
Surge
IEC/EN 61000-4-5
±0.5 kV, 1 kV
for line to line
±0.5kV, ±1 kV,
2 kV for line to
ground
±0.5 kV, 1 kV for
line to line
±0.5kV, ±1 kV, 2
kV for line to ground
The main power quality
should be a typical
commercial or hospital
environment.
Power frequency
(50 Hz / 60 Hz)
magnetic field
IEC/EN 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a specific
location in a typical
commercial or hospital
environment.