A1.5 Performance Specifications
Recording
Recorder
Thermal dot-matrix recorder
Printing Density
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25
mm/s)
Recorder Paper
Folded thermal paper: 210 mm×140 mm
Effective Width
210 mm
Paper Speed
5 mm/s, 6.25 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50
mm/s (±3%)
HR Calculator
HR Range
30 bpm~300 bpm
Accuracy
±1 bpm
ECG Unit
Filter
AC Filter: 50Hz / 60Hz / Off
DFT Filter: 0.01Hz / 0.05Hz / 0.32Hz / 0.67Hz
EMG Filter: 25Hz / 35Hz / 45Hz / Off
Low pass Filter: 350Hz / 300Hz / 270Hz / 150Hz / 100Hz /
75Hz
Leads Mode
9/12Lead
Analog-to-Digital
Converter
24 bits
Sampling
Frequency
64,000 /sec/channel
Time Constant
≥5s
Frequency
Response
0.01Hz ~ 350Hz(-3dB)
Gain
1.25, 2.5, 5, 10, 20, 10/5, AGC (mm/mV) (±5%)
Speed
5mm/s6.25mm/s10mm/s12.5mm/s25mm/s50mm/s
Input Impedance
≥100MΩ (10Hz)