User's Manual
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance
- 56 -
Put the dustproof coat on the main unit and sampling box after use and
prevent shaking them violently when moving to another place.
Prevent any liquid from seeping into the equipment; otherwise, the safety and
the performance of the electrocardiograph and sampling box cannot be
guaranteed.
ECG cable
The integrity of the ECG cable, including the main cable and lead wires,
should be checked regularly. Make sure that it is conductible.
Do not drag or twist the ECG cable with excessive stress. Hold the plug
instead of the cable when connecting or disconnecting the patient cable.
Align the ECG cable to avoid twisting, knotting, or crooking at a closed angle.
Store the lead wires appropriately to prevent any people from stumbling.
Once damage or aging of the ECG cable is found, replace it with a new one
immediately.
Reusable electrodes
1. Electrodes must be cleaned after use and ensure no gel is left on them.
2. Keep suction bulbs of chest electrodes away from sunshine and excessive
temperature.
3. After long-term use, the surfaces of electrodes will be oxidized because of
erosion and other causes. By this time, electrodes should be replaced to achieve
high data quality.
4. While usage will impact, it is expected that the electrocardiograph will be in
service for ten years. Replace the ECG cable and reusable electrodes according
to your actual use. It is recommended that you replace them once every year.
CAUTION
The device and accessories must be disposed of according to local regulations at the end
of their service lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal.