User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance
- 52 -
3. Remove the remaining disinfectant if appropriate with a dry cloth.
4. Air-dry the main unit/sampling box/ECG cable/reusable electrodes for at least 30
minutes in a ventilated and cool place.
The validated disinfectants for the main unit, sampling box, and ECG cables are:
Ethanol (75%)
Isopropanol (70%)
The validated disinfectant for the reusable electrodes is:
Isopropanol (70%)
If Ethanol or Isopropanol is used for both cleaning and disinfecting, a new cloth is
required for the disinfection step.
CAUTION
1. As for disinfection methods, do not use high-temperature, high-pressure vapor, or
ionizing radiation.
2. Do not use a chloric disinfectant such as chloride, sodium hypochlorite, etc.
3. Clean and disinfect reusable electrodes after each use.
WARNING
Turn off the power before disinfection. The mains supply must be switched off if it is used.
1. Switch off the main unit and disconnect it from the power cord.
2. Wipe the exterior surface of the electrocardiograph, patient cable, and reusable
electrodes (suction bulbs of chest electrodes and the clamps of limb electrodes)
using a soft cloth dampened with the disinfectant solution.
3. Wipe off the disinfectant solution with a dry cloth after disinfection if necessary.
4. Dry the electrocardiograph, patient cable, and reusable electrodes for at least 30
minutes in a ventilated and cool place.
11.4 Care and Maintenance
CAUTION
Operate the cardiograph, charge the battery, and store the battery at a temperature of
40 (104ºF ) or lower. Exposure to higher temperatures may reduce battery life, damage
the battery, and degrade overall cardiograph performance.
11.4.1 Recharge and Replacement of Battery