SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Error Messages
- 49 -
Chapter 10 Error Messages
Table 9-1 Error Messages and Causes
Message
Cause
Lead off
Electrodes fall off the patient, the patient
cable falls off the unit, or a high
polarization voltage occurs.
Network unavailable!
4G, WIFI, and Ethernet are off.
Network connection timeout! Please
check the network.
The network signal is poor, and the
network connection is failed.
Order server connection failed!
An IP address is wrong, or the server is off.
FTP connection failed!
An IP address is wrong, or the server is off.
Are you sure to stop sampling?
Tap during the acquisition of ECG data.
Overload
The direct current offset voltage on an
electrode is too high.
Inversion
Reversal of left and right arm electrodes?
No Paper
Recorder paper runs out or is not loaded.
Battery Weak
The battery is weak.
Transmitting fails
Data fails to be transmitted through
Ethernet or WIFI.
Memory Full
There is no space for saving more records.