User's Manual
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup
- 46 -
Item
Description
Axis calculation
method
Choose from: Area Method or Amplitude Method.
Tachycardia (greater
than)
Enter a heart rate. The default value is 100 bpm.
Bradycardia (less
than)
Enter a heart rate. The default value is 60 bpm.
P Wave Normal
Duration
Choose from: 110ms or 120ms.
When the P wave duration exceeds this setting, the system will
prompt you to P wave widen.
Algorithm Sensitivity
Choose from: Normal or Weak.
30us~2.0ms
Serious Diseases
Hint
Choose from: On, Off.
When On, diagnosis with severe illnesses will be marked with
“►◄” in the ECG report.
QTc Formulas
Choose from: Bazzet, Fridericia, Framingham, Hodges, and
QRS.
Bazzet, Fridericia, Framingham, Hodges, and QRS are
displayed in the ECG report as QTcBz, QTcFd, QTcFm, QTcHd,
and QTcQRS, respectively.
RV5+SV1
This measuring parameter will be displayed in auto
measurements, and ECG reports if selected. It is selected by
default.
RV6/SV2
This measuring parameter will be displayed in auto
measurements, and ECG reports if selected. It is not determined
by default.
RR/PP
This measuring parameter will be displayed in auto
measurements, and ECG reports if selected. This is not
determined by default.
Pacemaker
Choose from: Low, High.