Recording
Sequence
Choose from: Sequential, Simultaneous.
“Simultaneous” prints ECG data captured simultaneously for all
lead groups. “Sequential” prints ECG data captured at
sequential intervals for each lead group in turn.
Recording Device
Choose from: Thermal printer, WIFI printer, USB printer.
If a thermal printer is not configured, you can connect an
external printer.
Speed
Choose from: 5mm/s、6.25mm/s、10mm/s、12.5mm/s、25 mm/s、
50 mm/s。
Gain Setting
Choose from: 1.25 mm/mV, 2.5 mm/mV, 5 mm/mV, 10
mm/mV, 20 mm/mV, 10/5 mm/mV.
AGC
If selected, the gain will be automatically adjusted. The default
setting is Off.
Baseline Adjustment
Choose from: Horizontal, Auto, Off
Paper Marker
Choose from: Off, ON
9.3 Patient Information Setting
Item
Description
Patient Source ID
Choose from: Inpatient ID, Outpatient ID, Physical Examination
ID.
Patient ID
Choose from: Auto, Manual
Select Auto, and the system automatically generates a patient ID
according to the test date.
Select Manual, and you can manually enter a patient ID in the New
Patient window.
Default Patient ID
Enter four or fewer English letters or numbers as the prefix of
patient ID. The default setting is blank.
First/Last Name
If selected, first name and last name will be displayed in two
columns in the New Patient window.
If not selected, only the name is displayed in this window. It is the
default setting.
Age/Birthdate
Choose from: Manual, Birthdate
Select Manual, and you need to enter the patient's age in the