User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup
- 41 -
II for RHYT1, lead V1 for RHYT2, and lead V5 for RHYT3.
Reverse
Electrode Hint
If selected, the system prompts you if the electrodes are connected
in the opposite position. The default setting is On.
9.2 Print and Storage Setting
Item
Description
Print Report
Print ECG report after recording. The default setting is Off.
Auto Measure
If selected, the system provides you with auto measurements
after ECG recording. The default setting is On.
Auto diagnosis
If selected, the system provides you with auto diagnosis after
ECG recording. The default setting is On.
File Format
Choose from: SCP, FDA-XML, DICOM (Encapsulated PDF),
DICOM (ECG Waveform), PDF, JPG, BMP, PNG, and TIFF.
Determine the format of ECG files that are transmitted to the
server. If not selected, ECG files in DAT format will be
transferred.
Storage Location
Choose from: Local memory, External SD card.
External storage is
forbidden
Choose from: On, Off.
Report Format
Choose from:
Pediatric mode: 3x3, 6+3, 9x1
12-lead: 3x4+1R, 6x2+1R, 12x1.
NOTE:
In the case of an external printer, only the default report formats
are available.
Report Print
Channel
Define the channels to be printed in the ECG report.
The options are varied depending on the lead mode.
12-lead: 3 Channels, 6 Channels, 12 Channels, and Customize.
9-lead: 3 Channels, 6 Channels, 9 Channels, and Customize.
When you select “Customize”, you can choose the leads for 3 or
6 channels.