User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup
- 40 -
Chapter 9 System Setup
To set up the SE-1200 Pro&SE-1201 Pro ECG system, tap System Setting in the Main
Menu. Options underlined are default settings. You need to tap Save in each setting
window to save your changes if available.
9.1 Sampling Setting
Item
Description
Sampling Mode
Choose from: Real-time sampling or Pre-sampling
Select Pre-sample to initiate the acquisition of ECG data before
you tap . The latest 10 seconds of ECG data will be recorded.
Select a Real-time sample to initiate the acquisition of ECG data
after you tap .
Sampling Time
Set the duration for real-time acquisition, single-lead or three-lead
rhythm ECG acquisition. Enter a value between 10s and 1,800s.
The default is 10s for real-time acquisition, the 60s for single-lead
rhythm, and 20s for three-lead rhythm ECG.
NOTE: The pre-sampling time is fixed at 10s.
Intellectual
Sampling
If selected, the electrocardiograph automatically acquires ECGs
when it detects no poor ECG signals in 5 seconds. The default
setting is Off.
NOTE: This option applies to the Auto mode of standard ECG.
AI Pace
Detection
If selected, the system asks you if to display the pacing spike or not
when it detects a pacemaker pulse.
NOTE: This option applies to the Auto mode of standard ECG.
Enter the
Analysis screen
when sampling
finishes
If selected, the system automatically opens the ECG Analysis
screen when sampling finishes. The default setting is On.
Preview
If selected, the system opens the report preview page when the
ECG analysis finishes. The default setting is Off.
Lead Mode
Choose from: Pediatric mode, 12-Lead
Rhythm Leads
Select the leads to be recorded for rhythm strips. The default is lead