User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Editing ECG
- 36 -
Averages screen is opened as below.
1 Averages thumbnail
The averaged complexes of standard leads and ALL are displayed. Slide the thumbnail to
switch from one lead to another. Tap on one lead to display its magnified averages
2 Magnified averages
Six measurement lines are matched to six measurement points. They are P1 (onset of P
wave), P2 (offset of P wave), Q (onset of the QRS complex), S (offset of the QRS
complex), T1 (onset of T wave), and T2 (offset of T wave). Tap and slide a measurement
line of an averaged complex, and the same line of other averages will be automatically
moved. After moving these lines, tap Re-diagnose on the Averages screen to save
measurements. The measurements and diagnosis will be automatically updated.
Double-tap this portion, and a measurement pane will pop up. Slide the pane lines to get
real-time QRS duration, amplitude, and HR.
3 Detailed measurements of one lead.
Slide left to display the measurement results of all leads.
6.9 Comparing ECGs
SE-1200 Pro&SE-1201 Pro ECG allows you to compare ECGs with the same patient ID.
1. Open the Analysis screen.
2. Tap . Select Compare from the menu.
3. Select the ECGs you want to compare from the pop-up window.
4. Tap Compare. You can compare the ECG waveforms or averaged complexes.
1
3
2