3.2.3 Attaching the Disposable Electrodes
Disposable Electrode Connection (Clip Style)
1) Align all lead wires of the patient cable to avoid twisting, and connect the
clip/snap/banana socket adaptors to the patient cable.
2) Clean the electrode areas of the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the clip/snap/banana socket adaptors.
CAUTION
The disposable electrodes can only be used for one time.
3.3 Inspection Before Power-On
Before the operation, the following inspection procedures are recommended to avoid
safety hazards and get good ECG records.
1) Environment:
♦ Ensure no electromagnetic interference source around the equipment and
extensive medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment, etc. Turn off
these devices when necessary.
♦ Keep the examination room warm to avoid muscle tremor voltages in ECG
signals caused by cold.
2) Power Supply:
♦ If the mains supply is used, please check whether the power cord is connected
to the unit well. The grounded three-slot outlet should be used.
♦ When the battery capacity is low, recharge the battery before use.
3) Patient Cable:
♦ Ensure that the patient cable is connected to the unit firmly, and keep it far
away from the power cord.
4) Electrodes:
♦ Ensure that all electrodes are connected to the lead wires to the patient cable
correctly.
♦ Ensure that the chest electrodes do not in contact with each other.