User's Manual

SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance
- 20 -
3.2.3 Attaching the Disposable Electrodes
Disposable Electrode Connection (Clip Style)
1) Align all lead wires of the patient cable to avoid twisting, and connect the
clip/snap/banana socket adaptors to the patient cable.
2) Clean the electrode areas of the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the clip/snap/banana socket adaptors.
CAUTION
The disposable electrodes can only be used for one time.
3.3 Inspection Before Power-On
Before the operation, the following inspection procedures are recommended to avoid
safety hazards and get good ECG records.
1) Environment:
Ensure no electromagnetic interference source around the equipment and
extensive medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment, etc. Turn off
these devices when necessary.
Keep the examination room warm to avoid muscle tremor voltages in ECG
signals caused by cold.
2) Power Supply:
If the mains supply is used, please check whether the power cord is connected
to the unit well. The grounded three-slot outlet should be used.
When the battery capacity is low, recharge the battery before use.
3) Patient Cable:
Ensure that the patient cable is connected to the unit firmly, and keep it far
away from the power cord.
4) Electrodes:
Ensure that all electrodes are connected to the lead wires to the patient cable
correctly.
Ensure that the chest electrodes do not in contact with each other.