SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance
- 8 -
CAUTION
1. The access/operation of the electrocardiograph is restricted to authorized
personnel only. Assign only staff with a specific role the right to use the
electrocardiograph.
2. Ensure that all device components maintaining personal information (other
than removable media) are physically secure (i.e., cannot remove without
tools).
3. Users should make a backup of data periodically.
4. Ensure that the electrocardiograph is connected only to the device
authorized/approved by EDAN. Users should operate all EDAN deployed and
supported electrocardiographs within EDAN allowed specifications, including
approved software, software configuration, security configuration, etc.
5. Protect all the passwords to prevent unauthorized changes.
6. Before using a USB flash drive, anti-virus measures such as USB device virus
scanning should be carried out.
7. Firewalls and other security devices should be between the medical system
and any externally accessible systems. It’s recommended to use a Windows
defender firewall or any other firewall that can defend against Dos and DDoS
attacks and keep it up to date. Add the master computer software to the
whitelist.
8. Dos and DDoS protection of the router or switch must be turned on to defend
against attacks.
9. When building the networking environment: 1) If a wireless router is used,
please turn on the MAC address filtering function of the wireless router and
add the MAC address of the electrocardiograph to the rule list. The wireless
router only allows devices in the rule list to access the wireless network. 2) It is
suggested to build a VLAN, assign the LAN ports where the approved switch
port and electrocardiograph are in the same VLAN, and isolate it from other
VLANs.
10. Do not install in SE-1200 Pro&SE-1201 Pro ECG the other APPs or software that
the manufacturer does not configure.
11. When the device is returned for maintenance, disposed of, or removed from the
medical institution for other reasons, it is necessary to ensure that all patient
data are removed from the electrocardiograph.