SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual I
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual About this Manual P/N:01.54.459239 MPN:01.54.459239011 Release Date: July 2022 © Copyright EDAN INSTRUMENTS, INC. 2022. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It reminds us that the product shall be used strictly in complying with this manual. The user’s process failing to adhere to this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Terms Used in this Manual This guide is designed to give critical concepts on safety precautions. WARNING A WARNING label advises against specific actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides helpful information regarding a function or a procedure.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Indications for Use/Intended Use ....................................................................................... 1 1.2 Precautions for Use............................................................................................................. 1 1.3 Warnings and Cautions ..............
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual 6.6 Modifying Diagnosis ........................................................................................................ 35 6.7 Re-analysis ....................................................................................................................... 35 6.8 Averages View .................................................................................................................. 35 6.9 Comparing ECGs...............................
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance Chapter 1 Safety Guidance 1.1 Indications for Use/Intended Use The SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiographs are only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 1.3 Warnings and Cautions WARNING 1. The system is intended to use qualified physicians or personnel professionally trained. They should be familiar with the contents of this user manual before operation. 2. Only qualified service engineers can install this equipment, and only service engineers authorized by the manufacturer can open the shell. Otherwise, safety hazards may happen. 3.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance WARNING 10. Electrodes of dissimilar metals should not be used; otherwise, they may cause a high polarization voltage. 11. The disposable electrodes can only be used for one time. 12. Do not touch the patient, bed, table, or device while using the ECG and a defibrillator. 13. Do not touch accessible parts of electrical equipment and the patient simultaneously. 14. Disposable electrodes must be used during defibrillation. 15.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 20. SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied as a part of the system, to the multiple portable socket-outlet providing the system. 21. Do not connect any equipment or accessories that are not approved by the manufacturer or that are not IEC/EN 60601-1 approved in the system.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 29. Connecting any accessory (such as an external printer) or other devices (such as the computer) to this electrocardiograph makes a medical system.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 32. All the accessories connected to the system must be installed outside the patient vicinity if they do not meet the requirement of IEC/EN 60601-1. 33. The appliance coupler or mains plug is used as isolation means from supply mains. Position the electrocardiograph where the operator can easily access the disconnecting device. 34. The electrocardiograph should be placed on a flat surface or EDAN’s trolley.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 1.3.1 Protecting Personal Information Protecting personal health information is a significant component of security strategy. To protect personal information and ensure the proper device performance, the user should take necessary precautions by local laws and regulations and institutional policies.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance CAUTION 1. The access/operation of the electrocardiograph is restricted to authorized personnel only. Assign only staff with a specific role the right to use the electrocardiograph. 2. Ensure that all device components maintaining personal information (other than removable media) are physically secure (i.e., cannot remove without tools). 3. Users should make a backup of data periodically. 4.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance CAUTION 12. Please protect patients' information displayed on the screen and the data stored in the electrocardiograph. 13. The manufacturer or its authorized personnel can only upgrade the operating system of SE-1200 Pro&SE-1201 Pro ECG. If the user reinstalls SE-1200 Pro&SE-1201 Pro ECG with other operating systems not authorized by EDAN, the user shall bear the resulting security risks and losses.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 1.3.3 General Cautions CAUTION 1. Avoid liquid splash and excessive temperature. The temperature must be kept between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and 55 ºC during transportation and storage. 2. Do not use the equipment in a dusty environment with inadequate ventilation or in the presence of corrosive. 3.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance CAUTION 8. The medical electrical equipment needs to be installed and put into service according to Appendix 2 EMC Information. 9. The device and accessories must be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of them together with household garbage.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 1.4 List of Symbols No.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Safety Guidance 15 Dispose of method 16 Serial number General warning sign 17 (Background: Yellow; Symbol & Outline: Black) Refer to the instruction manual/booklet 18 (Background: Blue; Symbol: White) Non-ionizing 19* electromagnetic symbol 20 Environmentally friendly use period 21 This way up 22 Fragile, handle with care 23 Keep dry 24 Stacking limit by number 25 Handle with care 26 Do not step on 27 Front 28 radiation Co
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Operation Preparations Chapter 2 Product Information SE-1200 Pro&SE-1201 Pro ECG is new generation series of high-end smart electrocardiographs with a light body, which can meet the needs of multi-scenario and multi-service use.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Symbol Name Operation Preparations Description Mode To switch operation mode in the Resting ECG Test screen. To configure the system setting in the Resting ECG Test screen. To Setting exit, press this button again. Print/Stop Symbol ⚫ To start printing report(s). ⚫ To stop print when report(s) is being printed. ⚫ To stop paper feeding.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Operation Preparations Chapter 3 Operation Preparations 3.1 Preparing the Patient 3.1.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. It can decrease the patient's anxiety. Reassure the patient that the operation is painless. Privacy is critical for the patient to relax. When possible, prepare the patient in a quiet room or area where others can't see the patient.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Operation Preparations 3.2 Attaching Electrodes to the Patient WARNING 1. Make sure that all electrodes are connected to the patient correctly before operation. 2. Ensure that the conductive parts of electrodes and associated connectors, including neutral electrodes, do not contact the earth or any other conducting objects. NOTE: The quality of the ECG waveform will be affected by the contact resistance between the patient and the electrode.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Operation Preparations Electrodes Locations IEC AHA C1 V1 White/Red Brown/Red C2 V2 White/Yellow Brown/Yellow C3 V3 White/Green Brown/Green C4 V4 White/Brown Brown/Blue C5 V5 White/Black Brown/Orange C6 V6 White/Violet Brown/Violet L LA Yellow Black R RA Red White Fourth intercostal space at the right border of the sternum Fourth intercostal space at the left border of the sternum Fifth rib between C2 and C4 Fifth i
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual F LL Green Red N RL Black Green Operation Preparations Left leg Right leg 3.2.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance 3.2.3 Attaching the Disposable Electrodes Disposable Electrode Connection (Clip Style) 1) Align all lead wires of the patient cable to avoid twisting, and connect the clip/snap/banana socket adaptors to the patient cable. 2) Clean the electrode areas of the body surface with 75% alcohol. 3) Attach the disposable electrodes to the electrode positions on the body surface.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance 5) Patient: ♦ The patient should not come into contact with conducting objects such as earth, metal parts, etc. ♦ Ensure that the patient is warm and relaxed and breathes calmly. 3.4 Turning On/Off the Electrocardiograph WARNING 1. If the integrity of the external protective conductor is in doubt, the equipment should be powered by the battery. 2.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance CAUTION 3. If the electrocardiograph is turned off because of the low battery capacity or unexpected power failure, the settings or the current ECG report may not be saved. 4. The electrocardiograph cannot print an ECG report when the battery is weak. 5. The use of electrocardiograph accessories (such as barcode reader) will deplete battery power at a faster rate.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual 4. Basic Operation Guidance Paper feed. When you print an ECG report, if Paper Marker is set to On, the electrocardiograph can identify the black markers and stop at the crease of the paper for the convenience of tearing. NOTE: 1. When the folded thermal paper is used, the paper roller is unnecessary and must be taken out. 2. The two moving parts should be removed when using paper of 215mm in width. 3.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance Chapter 4 Basic Operation Guidance 4.1 Login If you have set a user name and login password, see the 8.7 user settings for details. After powering on, you need to enter the user name and password on the login interface. If you do not set up a user log-in, you will enter the reservation interface directly after powering on. 4.2 ECG recording The recording method varies depending on the type of ECG to be recorded.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance Start or stop ECG recording. NOTE: SE-1200 Pro&SE-1201Pro can record up to 1,800-second ECGs. Re-acquire ECGs. Auto Recording mode, including Auto and Manual. Tap on it to toggle between the two modes. In the Auto mode, the system will acquire ECGs at the set time and then print a report. If the user stops recording before the time, the system will print the last 10s of ECG.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance 4.3 Patient Tab The Patient tab displays all the patients that have ordered an ECG test. If patients’ priority is urgent, they will be listed at the top, and their information will be displayed in red. Button/Icon Description Return to the previous menu. Display the on-screen number of patients to be tested. This number updates in real-time.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance Enter new patient information. Delete patient information. Download batches of patient information from the PC. NOTE: A maximum of 200 latest orders can be downloaded from the server. Activate SE-1200 Pro&SE-1201Pro ECG’s infrared scanner to scan the barcode to retrieve the Outpatient / Inpatient / PE ID. Tap this icon again to stop scanning. Download the specific patient information from the PC.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance patient data will be loaded and displayed in the Patient tab. You have to configure the barcode before scanning to get the correct data. For information, see 7.3 Patient Information Setting. 4.4.3 Barcode scanner 1. Configure the barcode scanner. 2. For information about configuring the barcode scanner, please contact the manufacturer or your local distributor. 3.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Tap Basic Operation Guidance to download a batch of orders from the server. After download, patient data with the orders will be listed in the Patient tab. You can also slide down the tab for an update. NOTE: • To retrieve orders from the server, you need to purchase and install EDAN’s data management software on your PC. • Use the recommended barcode scanner. 4.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Basic Operation Guidance 2. Configure the setting. 3. Enter the sorting criteria, e.g., Outp./Inp./PE ID, name, request ID, etc. Tap OK. 4. Patient information that meets the criteria is displayed. To cancel sorting, tap Patient information before sorting is displayed. To queue patient information: 1. Tap in the Patient tab. Select Sequence Setting from the pop-up menu. 2. Configure the setting.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual File Management Chapter 5 ECG File Management Click the ECG File tab to manage the patient's ECG data. After the ECG data is recorded, it will be saved to the ECG File regardless of whether it is immediately analyzed. The ECG File of the SE-1200 Pro&SE-1201 Pro can store the data of 10s DAT examination with files larger than 10W. Button/Icon Description Search for ECGs that have been recorded.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual File Management Import ECGs from U disk. Export ECGs to the USB memory device. You can select the file format for export. See section 8.2 Print and Storage Setting. Upload ECGs to the server. You can choose one or more or even all ECGs to upload at a time. Delete ECGs. You can choose one or more or even all ECGs to delete at a time. Modify the patient information in an ECG file. You can modify all patient information except the ID and exam items.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Editing ECG Chapter 6 Editing ECG The system can be configured to analyze ECGs after recording automatically. To manually activate ECG data analysis, tap an ECG in the ECG File tab. Then the ECG Analysis screen is displayed. You can edit, measure, diagnose, and compare ECGs on this screen. The analysis screen is varied according to the ECG type. The following figure shows the analysis screen of standard ECG as an example.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Editing ECG Zoom in on an ECG To zoom in on the waveform, press and hold the waveform of a lead. The ECG waveform can be magnified up to 5 times by tapping ○ +. To exit magnification, tap on the X in the upper left corner. Manual Measurement To measure the ECG waveform, open the waveform magnification page. Tap on the ruler in the right lower corner. A measurement pane will be displayed. Tap on the ruler once more to exit the measurement.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Editing ECG For calculation of QTc, select System Setting > Parameter Setting > QTc Formulas. The default formula is Bazett. You can change the formula, but this takes effect only for the ECG analysis done after this change.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Editing ECG Averages screen is opened as below. 1 2 1 ○ 3 Averages thumbnail The averaged complexes of standard leads and ALL are displayed. Slide the thumbnail to switch from one lead to another. Tap on one lead to display its magnified averages 2 Magnified averages ○ Six measurement lines are matched to six measurement points.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Printing ECG Reports Chapter 7 Printing ECG Reports Tap on the Analysis screen, and select Print from the menu. After clicking print, the system will automatically print. Before printing for the first time, you need to set up the printer connection. The USB direct-connected printers are recommended to use HP 1112, HP2132, HP Desk Jet 4729,HP Desk Jet 3638,HP LaserJet Pro M202DW, HP M401,HP LaserJet P2035,HP 1020P,HP LaserJet Pro M403D.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Transmitting and Exporting ECGs Chapter 8 Transmitting and Exporting ECGs To transmit ECGs to server: 1. Log in to the FTP receiving software (including EDAN’s SE-1515) on your PC. 2. Turn on electrocardiograph. 3. Configure the transmission setup. 1) Find Ethernet Setting shortcut in the drop-down menu. 2) Turn on the Ethernet switch 3) Set Local IP, Gateway, and Subnet Mask. Or alternatively, select Auto Get IP. 4) Save Settings 4.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Transmitting and Exporting ECGs restored. To export ECGs to a USB storage device: 1. Connect a USB disk to the left side of electrocardiograph. 2. Open ECG File tab. 3. Tap . Select the ECGs you want to export. Or tap Select All. 4. Tap again to export the ECGs to the USB storage device.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup Chapter 9 System Setup To set up the SE-1200 Pro&SE-1201 Pro ECG system, tap System Setting in the Main Menu. Options underlined are default settings. You need to tap Save in each setting window to save your changes if available. 9.1 Sampling Setting Item Description Sampling Mode Choose from: Real-time sampling or Pre-sampling Select Pre-sample to initiate the acquisition of ECG data before you tap .
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup II for RHYT1, lead V1 for RHYT2, and lead V5 for RHYT3. Reverse If selected, the system prompts you if the electrodes are connected Electrode Hint in the opposite position. The default setting is On. 9.2 Print and Storage Setting Item Description Print Report Print ECG report after recording. The default setting is Off. Auto Measure If selected, the system provides you with auto measurements after ECG recording.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup Recording Choose from: Sequential, Simultaneous. Sequence “Simultaneous” prints ECG data captured simultaneously for all lead groups. “Sequential” prints ECG data captured at sequential intervals for each lead group in turn. Choose from: Thermal printer, WIFI printer, USB printer. Recording Device If a thermal printer is not configured, you can connect an external printer. Choose from: 5mm/s、6.25mm/s、10mm/s、12.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup textbox. Select Birthdate. You can slide the box to select the birth year, month, and day. The system will automatically calculate the age of the patient. Height/Weight BP Choose from: Off, cm/kg, inch/lb. Choose from: Off, mmHg, Kpa Hide patient data Choose from: Off, On viewable If On, patient name, age, and gender will be encrypted in the Patient tab and ECG File tab.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Barcode Setting System Setup Configure the start and end addresses, the male and female codes, an encoding mode, and then tap Save. NOTE: If not configured, the results of barcode scanning may be incorrect. 9.4 Filter Setting Item Description Baseline Filter ECG traces can contain some baseline wander caused by movement. You can reduce this wander of the baseline by applying a baseline filter.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Auto Upload System Setup Choose from: On, Off. When On, the system will automatically transmit ECG files to the server when the recording finishes. Delete file Choose from: On, Off. after uploading When On, the system will automatically delete ECG files from local memory when the files are automatically or manually transmitted to the server. Time Choose from: On, Off.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup Item Description Axis calculation Choose from: Area Method or Amplitude Method. method Tachycardia (greater Enter a heart rate. The default value is 100 bpm. than) Bradycardia (less Enter a heart rate. The default value is 60 bpm. than) P Wave Normal Choose from: 110ms or 120ms. Duration When the P wave duration exceeds this setting, the system will prompt you to P wave widen.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup Detection Rate Maximum Signal Choose from: ±5 mV, ±10 mV, ±20 mV. Range 9.7 User Setting You have to enter a password before configuring the user setting. Item Description User Login Choose from: On, Off. User login is selected by default. Add User Please enter the user’s real name, define their user name and password for login to SE-1200 Pro&SE-1201 Pro ECG.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual System Setup Function Activate DICOM, SCP, FDA-XML, communications. Select the one Activation you want to activate and enter the activation code. Communications here means retrieving orders from the server and uploading ECG data to the server. Set System Time Set or change the system time. Language Set the language according to your needs. 9.9 System Update Item Description Device Model Displays the model of the electrocardiograph.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Error Messages Chapter 10 Error Messages Table 9-1 Error Messages and Causes Message Cause Electrodes fall off the patient, the patient Lead off cable falls off the unit, or a high polarization voltage occurs. Network unavailable! 4G, WIFI, and Ethernet are off. Network connection timeout! Please The network signal is poor, and the check the network. network connection is failed.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance Chapter 11 Cleaning, Care, and Maintenance Use only the EDAN-approved substances and methods listed in this chapter to clean or disinfect your equipment. The warranty does not cover damage caused by using unapproved substances or methods. EDAN Instruments has validated the cleaning and disinfection instructions provided in this User Manual.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance required after each use. The validated cleaning agents for cleaning the electrocardiograph and ECG cable are: ⚫ Mild near-neutral detergent ⚫ Ethanol (75%) ⚫ Isopropanol (70%) The validated cleaning agent for cleaning the reusable electrodes is: ⚫ Mild near-neutral detergent Cleaning agents should be applied or removed using a clean, soft, non-abrasive cloth or paper towel.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance 3. Remove the remaining disinfectant if appropriate with a dry cloth. 4. Air-dry the main unit/sampling box/ECG cable/reusable electrodes for at least 30 minutes in a ventilated and cool place.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance 1) Capacity Identification The battery capacity can be identified according to the battery symbol in the top right corner of the LCD screen. 2) Recharge The electrocardiograph is equipped with a recharge control circuit together with the battery. When the unit is connected to the main supply, the battery will be recharged automatically.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance Consider a replacement if the battery has been fully charged and requires recharging after printing only a few ECGs. 11.4.2 Recorder Paper NOTE: Recorder paper provided by the manufacturer should be used. Other papers may shorten the life of the thermal print head. The deteriorated print head may lead to illegible ECG reports and block the advance of the paper.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance tests. a) Inspect the equipment and accessories for mechanical and functional damage. b) Inspect the safety-related labels for legibility. c) Inspect the fuse to verify compliance with the rated current and circuit-breaking characteristics. d) Verify that the device functions as described in the instructions for use. e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohms.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Cleaning, Care and Maintenance Put the dustproof coat on the main unit and sampling box after use and prevent shaking them violently when moving to another place. Prevent any liquid from seeping into the equipment; otherwise, the safety and the performance of the electrocardiograph and sampling box cannot be guaranteed. ECG cable The integrity of the ECG cable, including the main cable and lead wires, should be checked regularly.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Chapter 12 Accessories Table recommend11-1Standard Accessories List Accessory Part Number 01.13.036638 Power cord 01.13.037122 ECG Cable, Patient Cable (European) 01.57.471500 ECG Cable, Patient Cable (IEC) 01.57.471499 Adult Chest electrodes 01.57.040163 Adult Limb electrodes 01.57.040162 Recorder Paper 01.57.472248 Rechargeable Li-ion Battery, TWSLB-005 21.21.064149 AC-DC Adapter 21.21.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual U Disk 01.18.052245 Conductive Paste 01.25.078047 ECG Lead Adapter 01.57.471866 Dusting Cloths 01.57.471026 Pediatric Chest Electrodes 01.57.040168 Pediatric Limb Electrodes 01.57.040169 Electrode Adapters 01.57.040172 Snap/Banana Socket Adapters 01.57.471864 MT-201 Trolley 83.61.361191 Accessories 02.04.245388 CA-100 Lead Wire Bracket 02.04.245398 01.57.471858 01.57.471859 Disposable Electrodes 01.57.471860 01.57.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Warranty and Service Chapter 13 Warranty and Service 13.1 Warranty EDAN warrants that EDAN's products meet the labeled specifications of the products and will be free from defects in materials and quality that occur within the warranty period. The warranty is void in cases of: a) Damage caused by mishandling during shipping. b) Subsequent damage caused by improper use or maintenance.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Technical Specifications Appendix 1 Technical Specifications A1.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Technical Specifications A1.2 Environment Specifications Transport & Storage Working -20ºC (-4ºF)~+55ºC Temperature (+131ºF) Relative Humidity Atmospheric Pressure +5ºC (+41ºF) ~ +40ºC (+104ºF) 15%RH~95%RH 15%RH~95%RH Non-Condensing Non-Condensing 70kPa ~106kPa 70kPa ~106kPa A1.3 Physical Specifications Dimensions SE-1201Pro:298mm×258mm×82.5mm,±5mm Weight SE-1201Pro:4kg(excluding battery and recorder paper),±0.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Technical Specifications A1.5 Performance Specifications Recorder Thermal dot-matrix recorder 8 dots per mm / 200 dots per inch (amplitude axes) Printing Density 40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s) Recording Recorder Paper Folded thermal paper: 210 mm×140 mm Effective Width 210 mm 5 mm/s, 6.25 mm/s, 10 mm/s, 12.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Input Circuit Current Input Voltage Technical Specifications ≤0.01μA <±5mVp-p Range Calibration Voltage 1mV±1% DC Offset Voltage ±960mV, ±5% Noise ≤12.5 µV p-p Patient NC <10 µA (AC) / <10 µA (DC) SFC <50 µA (AC) / <50 µA (DC) NC <10 µA (AC) / <10 µA (DC) SFC <50 µA (AC) / <50 µA (DC) Leakage Current: Patient Auxiliary Current: ≥140dB(AC ON) CMRR ≥123dB(AC OFF) Amplitude ±500uV~±700mV Width 30us~2.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Technical Specifications ≤22 dBm (2.4GHz) Output Power ≤14 dBm (5GHz) 2.4GHz IEEE 802.11b: 1 Mbps to 11 Mbps DSSS IEEE 802.11g: 6 Mbps to 54 Mbps OFDM IEEE 802.11n: HT20 MCS7@17dB Transmit rate 5GHz IEEE 802.11a: 6 Mbps to 54 Mbps OFDM IEEE 802.11n: HT20 MCS0 to 7@16dB IEEE 802.11ac: VHT20 MCS8@15.5dB Data security Application-layer WEP,WPA/WPA2 PSK,802.1x EAP encryption No requirement. It's not used in real-time.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Appendix 2 EMC Information Electromagnetic Emissions Guidance and manufacturer’s declaration - electromagnetic emission The SE-1200 Pro&SE-1201 Pro ECG is intended for use in the electromagnetic environment specified below. The customer or the SE-1200 Pro&SE-1201 Pro ECG user should ensure that it is used in such an environment.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Electromagnetic Immunity Guidance and manufacturer’s declaration - electromagnetic immunity The SE-1200 Pro&SE-1201 Pro ECG is intended for use in the electromagnetic environment specified below. The customer or the SE-1200 Pro&SE-1201 Pro ECG user should ensure that it is used in such an environment.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Voltage dips, short interruptions, and voltage 0 % UT; 0.5 EMC Information 0 % UT; 0.5 cycle The main power input lines be a At 0°, 45°, 90°, 135°, should or At 0°, 45°, 90°, 180°, 225°, 270°, commercial environment.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Electromagnetic Immunity Guidance and manufacturer’s declaration - electromagnetic immunity The SE-1200 Pro&SE-1201 Pro ECG is intended for use in the electromagnetic environment specified below. The customer or the SE-1200 Pro&SE-1201 Pro ECG user should ensure that it is used in such an environment.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). As determined by an electromagnetic site survey, field strengths from fixed RF transmitters should be less than the compliance level in each frequency range.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Frequenc y (MHz) Maxim Brand a) (MHz) Service a) Modulation b) IMMUNITY um Distance TEST Power( (m) LEVEL W) (V/m) Pulse 385 380-390 TETRA 400 modulationb) 1.8 0.3 27 2 0.3 28 0.2 0.3 9 2 0.3 28 2 0.3 28 2 0.3 28 0.2 0.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual 5785 EMC Information 217 Hz Note: If necessary, to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The 1 m test distance is permitted by IEC/EN 61000-4-3. a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50% duty cycle square wave signal.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM Recommended separation distances between portable and mobile RF communications equipment and the SE-1200 Pro&SE-1201 Pro ECG The SE-1200 Pro&SE-1201 Pro ECG is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Abbreviations Appendix 3 Abbreviations Abbreviation Full Description LCD Liquid Crystal Display BP Blood Pressure ECG Electrocardiogram/Electrocardiograph HR Heart Rate AC Alternating Current USB Universal Serial Bus AGC Auto Gain Control NC Normal Condition SFC Single Fault Condition PDF Portable Document Format FDA-XML Food and Drug Administration-Extensible Markup Language DICOM Digital Imaging and Communications in Medicin
SE-1200 Pro&SE-1201 Pro Electrocardiograph User Manual Abbreviations Appendix 4 Installation instructions 1. The specific use scenario and installation setting diagram of the product. 2. This device is fixed used and can not move by Non-professionals. 3. This device must operate with a minim um distance of 20 cm between the radiator and user body.