User's Manual
SD5/SD6 Ultrasonic TableTop Doppler User Manual Ultrasound Intensity and Safety
- 42 -
Acoustic Output Reporting Table for Track 1
Non-autoscanning
Mode
Operating Mode: CW mode
Transducer: CD3.0
Working Frequency: 3.0MHz
Acoustic Output MI
I
SPTA.3
(mW/cm^2)
I
SPPA.3
(W/cm^2)
Global Maximum Value 0.01342 17.45 0.01745
Associated Acoustic
Parameter
P
r.3
(MPa) 0.02325
W
0
(mW) 43.90 43.90
f
c
(MHz) 3.00 3.00 3.00
Z
sp
(cm) 5.20 5.20 5.20
Beam dimensions
X
-
6
(cm) 0.6228 0.6228
Y
-
6
(cm) 1.372 1.372
PD (usec) N/A N/A
PRF (Hz) N/A N/A
EBD
Az.
(cm)
Φ1.7678
Ele.
(cm)
Φ1.7678
Operating Control
Conditions
Fixed