Product Information
SA Series Ambulatory Blood Pressure Monitor User Manual EMC Information
- 50 -
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture’s declaration - electromagnetic emission
The SA series device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SA series device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The SA series device is suitable for use in all
establishments, other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Not applicable
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
The SA series device is intended for use in the electromagnetic environment specified below.
The customer or the user of the SA SERIES DEVICE should assure that it is used in such an
environment.
Immunity test
IEC/EN 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.