Product Information
SA Series Ambulatory Blood Pressure Monitor User Manual Technical Specifications
- 49 -
Resolution
Blood pressure: 1 mmHg (0.133kPa)
Pulse: 1 bpm
Over-pressure protection
The cuff will be deflated when the measurement pressure
reaches 297±3mmHg (39.6±0.4 kPa).
Data memory
Flash memory stores up to 300 readings.
Sampling duration
24 hours (Time interval options are: 5, 10, 15, 20, 30, 45,
60, 90, 120 minutes)
Safety system
Maximum inflation pressure limited to 300 mmHg;
Maximum measurement time limited to 120 seconds
Bluetooth (configurable)
Transmitting Frequency
2400 MHz to 2483.5 MHz
Frequency Band
≤ 2MHz
Modulation Type
GFSK, DPSK, PSK
Transmitting Power
≤20 dBM (EIRP)
A1.6 Configuration Difference
The table below shows the configuration difference between models of ABPM.
Models
Configuration
Feature 1
Feature 2
Feature 3
Starting study in 5 min
Event recording
Study comparison
SA-10
SA-05
SA-06
SA-08
SA-09
Note: “” = Standard “” = Not Available