Product Information
SA Series Ambulatory Blood Pressure Monitor User Manual Technical Specifications
- 48 -
A1.2 Environment Specifications
Transport & Storage
Working
Temperature
-20ºC (-4ºF)~+55ºC (+131ºF)
+5ºC (+41ºF) ~ +40ºC (+104ºF)
Relative Humidity
15%RH~95%RH
Non-Condensing
15%RH~95%RH
Non-Condensing
Atmospheric Pressure
70 kPa ~106 kPa
70 kPa ~106 kPa
A1.3 Physical Specifications
Dimensions
106.5 mm × 62 mm × 27 mm, ±1 mm
Weight
160g ± 10g (excluding battery and cuff)
Display
0.96" screen
A1.4 Power Supply Specifications
Battery type
Two AA alkaline batteries
A1.5 Performance Specifications
ABP monitor
Method of measurement
Oscillometric
Static pressure range
0 to 300 mmHg (0 to 40 kPa), ±3 mmHg (±0.4 kPa)
Measurement pressure range
Systolic: 25 to 290 mmHg (3.3 to 38.6 kPa)
Diastolic: 10 to 250 mmHg (1.3 to 33.2 kPa)
Mean error: ±5 mmHg (±0.67 kPa)
Standard deviation: Max. 8 mmHg (1.07 kPa)
Pulse range
40 to 240 bpm, accurate within ±3 bpm or ±3.5%,
whichever is greater.