User's Manual

Acclarix LX9 Series Diagnostic Ultrasound System User Manual Measurements and Reports
- 113 -
PS/ED: measures the velocity at PS and ED points.
1. Invoke the generic measurement function.
2. Select Generic-> PS or ED.
3. Move the caliper to the PS or ED point on the strip.
4. Press the <Set> key to fix the point, and then PS or ED result displays.
RI: measures the velocity at PS and ED points, and calculates RI result.
1. Invoke the generic measurement function.
2. Select Generic-> RI.
3. Move the caliper to the PS point.
4. Press the <Set> key to fix the PS point.
5. Move the caliper to the ED point.
Presses of the <Update> key will toggle between active calipers, allowing you to re-position
the PS point or ED point.
6. Press the <Set> key to fix the ED point and complete the measurement.
S/D: measures the velocity at PS and ED points, and calculates S/D result.
1. Invoke the generic measurement function.
2. Select Generic-> PS, ED, RI, S/D.
3. Move the caliper to the PS point.
4. Press the <Set> key to fix the PS point.
5. Move the caliper to the ED point.
Presses of the <Update> key will toggle between active calipers, allowing you to re-position
the PS point or ED point.
6. Press the <Set> key to fix the ED point and complete the measurement.
8.1.3.2. Trace
The Trace measurement can provide a wide range of results, as shown below. The actual results
displayed will depend on Measure Setup(See section 10.4 for details).
PS
ED
MD
TAMax
PGmax
PGmean
PI
RI
S/D
VTI
Time
AT
DT
HR