User's Manual

- 220
-
Acclarix LX9 Series Diagnostic Ultrasound System User Manual
Specifications
Report Packages
Abdomen, Obstetrics, Small Parts, Vascular, Cardiac, Gynecology,
Urology, Pediatrics
Wi-Fi Specifications
Standard Conformance
IEEE802.11a/b/g/n/ac
Frequency Band
2.412-2.472GHz(@2.4GHz band), 5.180-5.825GHz(@5GHz band)
Modulation Technique
OFDM, CCK, DQPSK, BPSK
Typical Transmit Power
2dBm)
802.11b: 12.0 dBm
802.11g: 16.5dBm
802.11n: 16.5dBm
802.11a: 13.0dBm
802.11ac: 12.5dBm
Wi-Fi Quality of Service
Data rate
802.11b/g 1-54 Mbps
802.11a 6-54 Mbps
802.11n 1Stream MCS0-7
802.11n 2Stream MCS8-15
802.11ac MCS0-9
Data security
WPA/WPA2-PSK EAP encryption
Application-layer delay
No requirement. It's not used in real time.
Application-layer
reliability
No requirement. Application failure will be notified to the user
immediately.
System capacity
No more than one device will be allowed to connect with the
ultrasound system.
System anti-interference
Keep a distance from 2.4G band devices.
Network interruption
alarm
Network interruption is notified by disconnection icon and failure in
transmission is notified in Transfer Status window.
EMC test process
Wi-Fi function is not affected when the system is imposed with
radiation interference complied with IEC60601-1-2:2014 standard.
ECG Specifications
Patient Cable
IEC, AHA
Lead Mode
3-Lead: I, II, III
AC Filter
50 Hz, 60 Hz
HR Range and
Accuracy
Adult: 15 bpm to 300 bpm
Pediatric: 15 bpm to 350 bpm
Resolution: 1 bpm
Accuracy: ±1% or 1 bpm, whichever is greater.