User's Manual

Acclarix LX3 Series Diagnostic Ultrasound System User Manual Getting Started
- 31 -
3.3.4. Powering on/ off
Please review and follow the steps described in the Section
13.1.Daily Checklist
prior to powering on
the system.
To power on
1. Connect the system to a hospital grade power supply, or use the battery as the power supply.
2. Press the Power on/off key on the top left of the control panel.
To Login
If Password Protection is enabled (see section 11.1.7), the system displays a Login dialog when
booting up the system. Type in or select user name from the User Name drop-down list, then
enter password and click Login. At the first system powering on after you received the system, a
dialog will display for you to set the login password of the pre-defined administrator user.
For emergency use, click on Emergency to log on directly without entering user name and
password.
NOTE:
1. It is not suggested to use Emergency login under non-Emergency situation. Please login the
system with your account or switch to your account through the Switch User function (See
"To Switch Users" below) prior to imaging scanning.
2. Ensure the privacy of patient information and data created through Emergency login. It is
suggested to remove the patient exam data from the ultrasound system after Emergency
scanning for protecting patient data from being accessed by unauthorized users.
To Switch Users
If Password Protection is enabled, switching users is allowed without restarting the system.
1. Press Power on/off key, and the system will display a confirmation dialog box.
2. Select User from the confirmation dialog box. A login information dialog box will be displayed
providing access to change user.
3. Select another user from the User Name drop-down list and then enter password to login.
To power off:
1. Press the Power on/off key on the top left of control panel and the system displays a
confirmation dialog box.
2. Select “Shut Down” from the confirmation dialog box.
If the system is unresponsive, a long press of the Power on/off key will shut down the system directly.
NOTE:
1. Turn off and unplug the device after use.
2. Please unplug the AC plug from the power socket and disconnect the battery prior to storage.
Sleep mode
The system will enter a sleep mode that maintains exam information while using minimal power.
There are three events that can invoke sleep mode:
No user input for a configurable amount of time. Please see System Set-up to configure this
time.